Back to search
FDAFDA · Food and drugs·Recall number Z-1487-2024

Bed Assist Bars, labeled as follows: a) Medline, Model Number BA6800, b) Medline Martha Stewart, Model Number MDS6800BAMMS

Published April 17, 2024 · Recall initiated: March 5, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medline is recalling certain lots of our Adult Portable Bed Rails that were manufactured after the mandatory compliance requirement (ASTM F3186-17, Standard Specification for Adult Portable Bed Rails and Related Products and 16 CFR Part 1270, Safety Standard for Adult Portable Bed Rails) and failed to meet all mandatory requirements outlined in these standards.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
Distributor
Worldwide distribution - US Nationwide and the country of Canada.
Lot
a) Medline, Model Number BA6800, Lot Numbers: 88523090002, 88523090003, 88523100001, 88523100002, 88523100003, 88523110002, 88523110003, 88523120002, 88523120003, 88524010002, 88524020001, 88524020002, 88524030001; b) Medline Martha Stewart, Model Number MDS6800BAMMS, Lot Numbers: 39323090001, 39323090002, 39323100001, 39323100002, 39323110001, 39323110002, 39323120001, 39323120002, 39324010001, 39324010002;
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card