Back to search
FDAFDA · Food and drugs·Recall number Z-1488-2024

Bed Assist Bars, labeled as follows: a) Medline SwivAssist Bar, Model Number MDS6800SWIV (case of 3), b) Medline SwivAssist Bar, Model Number MDS6800SWIVRH (individual)

Published April 17, 2024 · Recall initiated: March 5, 2024

Summary

FDA records medical device Class II recall Z-1488-2024, published April 17, 2024 and initiated March 5, 2024. Recalling firm: MEDLINE INDUSTRIES, LP - Northfield; Status: Ongoing; Reported quantity: 79 devices; Lot: a) Medline SwivAssist Bar, Model Number MDS6800SWIV (case of 3), Lot Numbers: 15123110004, 15123110005, 15124010001, 15124020002; b) Medline SwivAssist Bar, Model Number MDS6800SWIVRH, Lot Numbers: 15123110004, 15123110005, 15123120001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medline is recalling certain lots of our Adult Portable Bed Rails that were manufactured after the mandatory compliance requirement (ASTM F3186-17, Standard Specification for Adult Portable Bed Rails and Related Products and 16 CFR Part 1270, Safety Standard for Adult Portable Bed Rails) and failed to meet all mandatory requirements outlined in these standards.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
Distributor
Worldwide distribution - US Nationwide and the country of Canada.
Lot
a) Medline SwivAssist Bar, Model Number MDS6800SWIV (case of 3), Lot Numbers: 15123110004, 15123110005, 15124010001, 15124020002; b) Medline SwivAssist Bar, Model Number MDS6800SWIVRH, Lot Numbers: 15123110004, 15123110005, 15123120001
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Bed Assist Bars, labeled as follows: a) Medline SwivAssist Bar, Model Number MDS6800SWIV (case of 3), b) Medline SwivAssist Bar, Model Number MDS6800SWIVRH (individual) recalled?

Yes. FDA published recall Z-1488-2024 on April 17, 2024. Reported hazard: Medline is recalling certain lots of our Adult Portable Bed Rails that were manufactured after the mandatory compliance requirement (ASTM F3186-17, Standard Specification for Adult Portable Bed Rails and Related Products and 16 CFR Part 1270, Safety Standard for Adult Portable Bed Rails) and failed to meet all mandatory requirements outlined in these standards.

What should I do if I have this product?

Voluntary: Firm initiated · Letter