Back to search
FDAFDA · Food and drugs·Recall number Z-1490-2012

Sigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification Name: Infusion Pump -- Class 2 medical device -- SIGMA LLC, Medina, NY 14103 Product Usage: The Spectrum Infusion Pump is intended to be used for the controlled administration of intravenous fluids. These fluids may include blood, blood products or mixtures of pharmaceutical drugs for required patient therapy. The Spectrum is used in conjunction with legally marked intravenous administration sets and medications provided by the user. The Master Drug Library is a software package that will add additional features to the Spectrum Infusion Pump. The Master Drug Library will permit electronic communications with the Spectrum pump and other external peripheral devices. The intended use of the Spectrum pump includes common drug error prevention, through the stand alone settings features of the pump. This includes drug parameter limits and associated drug name identification. With the Master Drug Library, the intended use is to reduce user errors associated with drug selection, drug dose rates, drug dose concentrations, and patient identification associated with the prescribed drug.

Published June 20, 2012 · Recall initiated: July 6, 2011

Summary

FDA records medical device Class I recall Z-1490-2012, published June 20, 2012 and initiated July 6, 2011. Recalling firm: Baxter Healthcare Corporation; Status: Terminated; Reported quantity: 63,225 pumps total (corrected number): Domestic - 63,032 pumps; Canada - 29 pumps; Okinawa - 164 pumps; Lot: Serial numbers ranging from 700000 through 794213: 700000, 700001, 700002, 700003, 700004, 700005, 700006, 700007, 700010, 700012, 700013, 700014, 700017, 700018, 700020, 700021, 700022, 700023, 700024, 700029, 700030, 700031, 700032, 700033, 700034, 700035, 700037, 700038, 700039, 700040, 700041, 700042, 700043, 700045, 700046, 700047, 700048, 700049, 700050, 700051, 700052, 700053, 700054, 700055, 700057, 700058, 700059, 700060, 700061, 700062, 700063, 700064, 700065, 700066, 700067, 700069, 700071, 700073, 700075, 700076, 700077, 700080, 700081, 700082, 700085, 700086, 700088, 700089, 700090, 700093, 700094, 700095, 700097, 700098, 700100, 700101, 700103, 700104, 700106, 700108, 700109, 700110, 700111, 700112, 700114, 700117, 700118, 700119, 700120, 700121, 700123, 700126, 700127, 700129, 700130, 700132, 700133, 700134, 700135, 700137, 700138, 700139, 700141, 700142, 700143, 700144, 700145, 700147, 700149, 700151, 700152, 700153, 700156, 700157, 700158, 700159, 700161, 700162, 700163, 700165, 700166, 700168, 700169, 700170, 700172, 700173, 700174, 700175, 700176, 700177, 700178, 700179, 700180, 700183, 700184, 700187, 700189, 700193, 700194, 700195, 700197, 700201, 700202, 700203, 700204, 700205, 700206, 700207, 700208, 700209, 700210, 700211, 700212, 700213, 700214, 700215, 700216, 700217, 700218, 700219, 700220, 700222, 700223, 700224, 700225, 700226, 700229, 700231, 700232, 700233, 700234, 700235, 700236, 700237, 700238, 700239, 700240, 700241, 700242, 700243, 700244, 700245, 700246, 700248, 700249, 700250, 700251, 700252, 700253, 700255, 700256, 700257, 700258, 700259, 700260, 700263, 700265, 700267, 700268, 700269, 700271, 700272, 700273, 700274, 700275, 700277, 700278, 700280, 700282, 700283, 700284, 700285, 700286, 700287, 700288, 700289, 700292, 700293, 700294, 700295, 700297, 700298, 700299, 700301, 700302, 700303, 700305, 700306, 700307, 700308, 700309, 700310, 700311, 700312, 700314, 700316, 700317, 700318, 700319, 700320, 700321, 700322, 700323, 700324, 700326, 700327, 700328, 700329, 700330, 700331, 700332, 700333, 700334, 700335, 700338, 700339, 700340, 700341, 700343, 700344, 700345, 700346, 700348, 700349, 700350, 700351, 700352, 700353, 700354, 700355, 700357, 700358, 700359, 700360, 700361, 700362, 700363, 700364, 700365, 700366, 700367, 700368, 700369, 700370, 700371, 700372, 700373, 700374, 700375, 700376, 700377, 700378, 700379, 700380, 700381, 700382, 700383, 700384, 700385, 700386, 700387, 700388, 700389, 700390, 700392, 700395, 700396, 700398, 700399, 700400, 700401, 700402, 700403, 700404, 700405, 700406, 700407, 700408, 700409, 700410, 700411, 700412, 700413, 700414, 700415, 700416, 700417, 700419, 700420, 700421, 700422, 700423, 700424, 700425, 700426, 700427, 700428, 700429, 700433, 700434, 700436, 700437, 700438, 700439, 700440, 700441, 700442, 700443, 700444, 700446, 700447, 700448, 700449, 700450, 700451, 700452, 700453, 700454, 700455, 700456, 700457, 700458, 700459, 700460, 700461, 700462, 700463, 700464, 700466, 700468, 700470, 700471, 700472, 700473, 700474, 700475, 700476, 700478, 700479, 700480, 700481, 700483, 700484, 700485, 700489, 700490, 700491, 700494, 700495, 700496, 700497, 700502, 700503, 700504, 700505, 700506, 700507, 700508, 700509, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700517, 700519, 700520, 700524, 700525, 700526, 700527, 700528, 700530, 700531, 700532, 700533, 700534, 700536, 700537, 700538, 700539, 700540, 700542, 700548, 700552, 700553, 700555, 700556, 700557, 700558, 700560, 700561, 700562, 700563, 700564, 700565, 700566, 700567, 700568, 700569, 700570, 700571, 700572, 700573, 700574, 700575, 700576, 700577, 700578, 700579, 700580, 700581, 700582, 700583, 700584, 700585, 700586, 700587, 700588, 700589, 700590, 700591, 700592, 700593, 700594, 700595, 700596, 700597, 700598, 700599, 700600, 700601, 700602, 700603, 700604, 700605, 700606, 700607, 700608, 700611, 700612, 700613, 700614, 700616, 700617, 700618, 700619, 700620, 700621, 700622, 70062. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Baxter Healthcare Corporation
Distributor
Worldwide Distribution - USA Nationwide including Puerto Rico and the countries of Canada and Okinawa
Lot
Serial numbers ranging from 700000 through 794213: 700000, 700001, 700002, 700003, 700004, 700005, 700006, 700007, 700010, 700012, 700013, 700014, 700017, 700018, 700020, 700021, 700022, 700023, 700024, 700029, 700030, 700031, 700032, 700033, 700034, 700035, 700037, 700038, 700039, 700040, 700041, 700042, 700043, 700045, 700046, 700047, 700048, 700049, 700050, 700051, 700052, 700053, 700054, 700055, 700057, 700058, 700059, 700060, 700061, 700062, 700063, 700064, 700065, 700066, 700067, 700069, 700071, 700073, 700075, 700076, 700077, 700080, 700081, 700082, 700085, 700086, 700088, 700089, 700090, 700093, 700094, 700095, 700097, 700098, 700100, 700101, 700103, 700104, 700106, 700108, 700109, 700110, 700111, 700112, 700114, 700117, 700118, 700119, 700120, 700121, 700123, 700126, 700127, 700129, 700130, 700132, 700133, 700134, 700135, 700137, 700138, 700139, 700141, 700142, 700143, 700144, 700145, 700147, 700149, 700151, 700152, 700153, 700156, 700157, 700158, 700159, 700161, 700162, 700163, 700165, 700166, 700168, 700169, 700170, 700172, 700173, 700174, 700175, 700176, 700177, 700178, 700179, 700180, 700183, 700184, 700187, 700189, 700193, 700194, 700195, 700197, 700201, 700202, 700203, 700204, 700205, 700206, 700207, 700208, 700209, 700210, 700211, 700212, 700213, 700214, 700215, 700216, 700217, 700218, 700219, 700220, 700222, 700223, 700224, 700225, 700226, 700229, 700231, 700232, 700233, 700234, 700235, 700236, 700237, 700238, 700239, 700240, 700241, 700242, 700243, 700244, 700245, 700246, 700248, 700249, 700250, 700251, 700252, 700253, 700255, 700256, 700257, 700258, 700259, 700260, 700263, 700265, 700267, 700268, 700269, 700271, 700272, 700273, 700274, 700275, 700277, 700278, 700280, 700282, 700283, 700284, 700285, 700286, 700287, 700288, 700289, 700292, 700293, 700294, 700295, 700297, 700298, 700299, 700301, 700302, 700303, 700305, 700306, 700307, 700308, 700309, 700310, 700311, 700312, 700314, 700316, 700317, 700318, 700319, 700320, 700321, 700322, 700323, 700324, 700326, 700327, 700328, 700329, 700330, 700331, 700332, 700333, 700334, 700335, 700338, 700339, 700340, 700341, 700343, 700344, 700345, 700346, 700348, 700349, 700350, 700351, 700352, 700353, 700354, 700355, 700357, 700358, 700359, 700360, 700361, 700362, 700363, 700364, 700365, 700366, 700367, 700368, 700369, 700370, 700371, 700372, 700373, 700374, 700375, 700376, 700377, 700378, 700379, 700380, 700381, 700382, 700383, 700384, 700385, 700386, 700387, 700388, 700389, 700390, 700392, 700395, 700396, 700398, 700399, 700400, 700401, 700402, 700403, 700404, 700405, 700406, 700407, 700408, 700409, 700410, 700411, 700412, 700413, 700414, 700415, 700416, 700417, 700419, 700420, 700421, 700422, 700423, 700424, 700425, 700426, 700427, 700428, 700429, 700433, 700434, 700436, 700437, 700438, 700439, 700440, 700441, 700442, 700443, 700444, 700446, 700447, 700448, 700449, 700450, 700451, 700452, 700453, 700454, 700455, 700456, 700457, 700458, 700459, 700460, 700461, 700462, 700463, 700464, 700466, 700468, 700470, 700471, 700472, 700473, 700474, 700475, 700476, 700478, 700479, 700480, 700481, 700483, 700484, 700485, 700489, 700490, 700491, 700494, 700495, 700496, 700497, 700502, 700503, 700504, 700505, 700506, 700507, 700508, 700509, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700517, 700519, 700520, 700524, 700525, 700526, 700527, 700528, 700530, 700531, 700532, 700533, 700534, 700536, 700537, 700538, 700539, 700540, 700542, 700548, 700552, 700553, 700555, 700556, 700557, 700558, 700560, 700561, 700562, 700563, 700564, 700565, 700566, 700567, 700568, 700569, 700570, 700571, 700572, 700573, 700574, 700575, 700576, 700577, 700578, 700579, 700580, 700581, 700582, 700583, 700584, 700585, 700586, 700587, 700588, 700589, 700590, 700591, 700592, 700593, 700594, 700595, 700596, 700597, 700598, 700599, 700600, 700601, 700602, 700603, 700604, 700605, 700606, 700607, 700608, 700611, 700612, 700613, 700614, 700616, 700617, 700618, 700619, 700620, 700621, 700622, 70062
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Sigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification Name: Infusion Pump -- Class 2 medical device -- SIGMA LLC, Medina, NY 14103 Product Usage: The Spectrum Infusion Pump is intended to be used for the controlled administration of intravenous fluids. These fluids may include blood, blood products or mixtures of pharmaceutical drugs for required patient therapy. The Spectrum is used in conjunction with legally marked intravenous administration sets and medications provided by the user. The Master Drug Library is a software package that will add additional features to the Spectrum Infusion Pump. The Master Drug Library will permit electronic communications with the Spectrum pump and other external peripheral devices. The intended use of the Spectrum pump includes common drug error prevention, through the stand alone settings features of the pump. This includes drug parameter limits and associated drug name identification. With the Master Drug Library, the intended use is to reduce user errors associated with drug selection, drug dose rates, drug dose concentrations, and patient identification associated with the prescribed drug. recalled?

Yes. FDA published recall Z-1490-2012 on June 20, 2012. Reported hazard: Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.

What should I do if I have this product?

Voluntary: Firm initiated · Letter