Sigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification Name: Infusion Pump -- Class 2 medical device -- SIGMA LLC, Medina, NY 14103 Product Usage: The Spectrum Infusion Pump is intended to be used for the controlled administration of intravenous fluids. These fluids may include blood, blood products or mixtures of pharmaceutical drugs for required patient therapy. The Spectrum is used in conjunction with legally marked intravenous administration sets and medications provided by the user. The Master Drug Library is a software package that will add additional features to the Spectrum Infusion Pump. The Master Drug Library will permit electronic communications with the Spectrum pump and other external peripheral devices. The intended use of the Spectrum pump includes common drug error prevention, through the stand alone settings features of the pump. This includes drug parameter limits and associated drug name identification. With the Master Drug Library, the intended use is to reduce user errors associated with drug selection, drug dose rates, drug dose concentrations, and patient identification associated with the prescribed drug.
Published June 20, 2012 · Recall initiated: July 6, 2011
Summary
FDA records medical device Class I recall Z-1490-2012, published June 20, 2012 and initiated July 6, 2011. Recalling firm: Baxter Healthcare Corporation; Status: Terminated; Reported quantity: 63,225 pumps total (corrected number): Domestic - 63,032 pumps; Canada - 29 pumps; Okinawa - 164 pumps; Lot: Serial numbers ranging from 700000 through 794213: 700000, 700001, 700002, 700003, 700004, 700005, 700006, 700007, 700010, 700012, 700013, 700014, 700017, 700018, 700020, 700021, 700022, 700023, 700024, 700029, 700030, 700031, 700032, 700033, 700034, 700035, 700037, 700038, 700039, 700040, 700041, 700042, 700043, 700045, 700046, 700047, 700048, 700049, 700050, 700051, 700052, 700053, 700054, 700055, 700057, 700058, 700059, 700060, 700061, 700062, 700063, 700064, 700065, 700066, 700067, 700069, 700071, 700073, 700075, 700076, 700077, 700080, 700081, 700082, 700085, 700086, 700088, 700089, 700090, 700093, 700094, 700095, 700097, 700098, 700100, 700101, 700103, 700104, 700106, 700108, 700109, 700110, 700111, 700112, 700114, 700117, 700118, 700119, 700120, 700121, 700123, 700126, 700127, 700129, 700130, 700132, 700133, 700134, 700135, 700137, 700138, 700139, 700141, 700142, 700143, 700144, 700145, 700147, 700149, 700151, 700152, 700153, 700156, 700157, 700158, 700159, 700161, 700162, 700163, 700165, 700166, 700168, 700169, 700170, 700172, 700173, 700174, 700175, 700176, 700177, 700178, 700179, 700180, 700183, 700184, 700187, 700189, 700193, 700194, 700195, 700197, 700201, 700202, 700203, 700204, 700205, 700206, 700207, 700208, 700209, 700210, 700211, 700212, 700213, 700214, 700215, 700216, 700217, 700218, 700219, 700220, 700222, 700223, 700224, 700225, 700226, 700229, 700231, 700232, 700233, 700234, 700235, 700236, 700237, 700238, 700239, 700240, 700241, 700242, 700243, 700244, 700245, 700246, 700248, 700249, 700250, 700251, 700252, 700253, 700255, 700256, 700257, 700258, 700259, 700260, 700263, 700265, 700267, 700268, 700269, 700271, 700272, 700273, 700274, 700275, 700277, 700278, 700280, 700282, 700283, 700284, 700285, 700286, 700287, 700288, 700289, 700292, 700293, 700294, 700295, 700297, 700298, 700299, 700301, 700302, 700303, 700305, 700306, 700307, 700308, 700309, 700310, 700311, 700312, 700314, 700316, 700317, 700318, 700319, 700320, 700321, 700322, 700323, 700324, 700326, 700327, 700328, 700329, 700330, 700331, 700332, 700333, 700334, 700335, 700338, 700339, 700340, 700341, 700343, 700344, 700345, 700346, 700348, 700349, 700350, 700351, 700352, 700353, 700354, 700355, 700357, 700358, 700359, 700360, 700361, 700362, 700363, 700364, 700365, 700366, 700367, 700368, 700369, 700370, 700371, 700372, 700373, 700374, 700375, 700376, 700377, 700378, 700379, 700380, 700381, 700382, 700383, 700384, 700385, 700386, 700387, 700388, 700389, 700390, 700392, 700395, 700396, 700398, 700399, 700400, 700401, 700402, 700403, 700404, 700405, 700406, 700407, 700408, 700409, 700410, 700411, 700412, 700413, 700414, 700415, 700416, 700417, 700419, 700420, 700421, 700422, 700423, 700424, 700425, 700426, 700427, 700428, 700429, 700433, 700434, 700436, 700437, 700438, 700439, 700440, 700441, 700442, 700443, 700444, 700446, 700447, 700448, 700449, 700450, 700451, 700452, 700453, 700454, 700455, 700456, 700457, 700458, 700459, 700460, 700461, 700462, 700463, 700464, 700466, 700468, 700470, 700471, 700472, 700473, 700474, 700475, 700476, 700478, 700479, 700480, 700481, 700483, 700484, 700485, 700489, 700490, 700491, 700494, 700495, 700496, 700497, 700502, 700503, 700504, 700505, 700506, 700507, 700508, 700509, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700517, 700519, 700520, 700524, 700525, 700526, 700527, 700528, 700530, 700531, 700532, 700533, 700534, 700536, 700537, 700538, 700539, 700540, 700542, 700548, 700552, 700553, 700555, 700556, 700557, 700558, 700560, 700561, 700562, 700563, 700564, 700565, 700566, 700567, 700568, 700569, 700570, 700571, 700572, 700573, 700574, 700575, 700576, 700577, 700578, 700579, 700580, 700581, 700582, 700583, 700584, 700585, 700586, 700587, 700588, 700589, 700590, 700591, 700592, 700593, 700594, 700595, 700596, 700597, 700598, 700599, 700600, 700601, 700602, 700603, 700604, 700605, 700606, 700607, 700608, 700611, 700612, 700613, 700614, 700616, 700617, 700618, 700619, 700620, 700621, 700622, 70062. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- Worldwide Distribution - USA Nationwide including Puerto Rico and the countries of Canada and Okinawa
- Lot
- Serial numbers ranging from 700000 through 794213: 700000, 700001, 700002, 700003, 700004, 700005, 700006, 700007, 700010, 700012, 700013, 700014, 700017, 700018, 700020, 700021, 700022, 700023, 700024, 700029, 700030, 700031, 700032, 700033, 700034, 700035, 700037, 700038, 700039, 700040, 700041, 700042, 700043, 700045, 700046, 700047, 700048, 700049, 700050, 700051, 700052, 700053, 700054, 700055, 700057, 700058, 700059, 700060, 700061, 700062, 700063, 700064, 700065, 700066, 700067, 700069, 700071, 700073, 700075, 700076, 700077, 700080, 700081, 700082, 700085, 700086, 700088, 700089, 700090, 700093, 700094, 700095, 700097, 700098, 700100, 700101, 700103, 700104, 700106, 700108, 700109, 700110, 700111, 700112, 700114, 700117, 700118, 700119, 700120, 700121, 700123, 700126, 700127, 700129, 700130, 700132, 700133, 700134, 700135, 700137, 700138, 700139, 700141, 700142, 700143, 700144, 700145, 700147, 700149, 700151, 700152, 700153, 700156, 700157, 700158, 700159, 700161, 700162, 700163, 700165, 700166, 700168, 700169, 700170, 700172, 700173, 700174, 700175, 700176, 700177, 700178, 700179, 700180, 700183, 700184, 700187, 700189, 700193, 700194, 700195, 700197, 700201, 700202, 700203, 700204, 700205, 700206, 700207, 700208, 700209, 700210, 700211, 700212, 700213, 700214, 700215, 700216, 700217, 700218, 700219, 700220, 700222, 700223, 700224, 700225, 700226, 700229, 700231, 700232, 700233, 700234, 700235, 700236, 700237, 700238, 700239, 700240, 700241, 700242, 700243, 700244, 700245, 700246, 700248, 700249, 700250, 700251, 700252, 700253, 700255, 700256, 700257, 700258, 700259, 700260, 700263, 700265, 700267, 700268, 700269, 700271, 700272, 700273, 700274, 700275, 700277, 700278, 700280, 700282, 700283, 700284, 700285, 700286, 700287, 700288, 700289, 700292, 700293, 700294, 700295, 700297, 700298, 700299, 700301, 700302, 700303, 700305, 700306, 700307, 700308, 700309, 700310, 700311, 700312, 700314, 700316, 700317, 700318, 700319, 700320, 700321, 700322, 700323, 700324, 700326, 700327, 700328, 700329, 700330, 700331, 700332, 700333, 700334, 700335, 700338, 700339, 700340, 700341, 700343, 700344, 700345, 700346, 700348, 700349, 700350, 700351, 700352, 700353, 700354, 700355, 700357, 700358, 700359, 700360, 700361, 700362, 700363, 700364, 700365, 700366, 700367, 700368, 700369, 700370, 700371, 700372, 700373, 700374, 700375, 700376, 700377, 700378, 700379, 700380, 700381, 700382, 700383, 700384, 700385, 700386, 700387, 700388, 700389, 700390, 700392, 700395, 700396, 700398, 700399, 700400, 700401, 700402, 700403, 700404, 700405, 700406, 700407, 700408, 700409, 700410, 700411, 700412, 700413, 700414, 700415, 700416, 700417, 700419, 700420, 700421, 700422, 700423, 700424, 700425, 700426, 700427, 700428, 700429, 700433, 700434, 700436, 700437, 700438, 700439, 700440, 700441, 700442, 700443, 700444, 700446, 700447, 700448, 700449, 700450, 700451, 700452, 700453, 700454, 700455, 700456, 700457, 700458, 700459, 700460, 700461, 700462, 700463, 700464, 700466, 700468, 700470, 700471, 700472, 700473, 700474, 700475, 700476, 700478, 700479, 700480, 700481, 700483, 700484, 700485, 700489, 700490, 700491, 700494, 700495, 700496, 700497, 700502, 700503, 700504, 700505, 700506, 700507, 700508, 700509, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700517, 700519, 700520, 700524, 700525, 700526, 700527, 700528, 700530, 700531, 700532, 700533, 700534, 700536, 700537, 700538, 700539, 700540, 700542, 700548, 700552, 700553, 700555, 700556, 700557, 700558, 700560, 700561, 700562, 700563, 700564, 700565, 700566, 700567, 700568, 700569, 700570, 700571, 700572, 700573, 700574, 700575, 700576, 700577, 700578, 700579, 700580, 700581, 700582, 700583, 700584, 700585, 700586, 700587, 700588, 700589, 700590, 700591, 700592, 700593, 700594, 700595, 700596, 700597, 700598, 700599, 700600, 700601, 700602, 700603, 700604, 700605, 700606, 700607, 700608, 700611, 700612, 700613, 700614, 700616, 700617, 700618, 700619, 700620, 700621, 700622, 70062
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATruSystem 7500 surgical table, Model/Catalog Number: 4091000November 11, 2026
- FDATruSystem 7000 U Operating Table (dV), Product Code 1723633November 11, 2026
- FDATruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087November 11, 2026
- FDAOperating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788November 11, 2026
- FDAOrchids brand Smoked Oysters in cottonseed oil, salt added. Net Wt. 3 oz Ingredients: oysters, salt, water.June 20, 2012
- FDAPhilips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- NationwideJune 20, 2012
Questions about this recall
Was Sigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification Name: Infusion Pump -- Class 2 medical device -- SIGMA LLC, Medina, NY 14103 Product Usage: The Spectrum Infusion Pump is intended to be used for the controlled administration of intravenous fluids. These fluids may include blood, blood products or mixtures of pharmaceutical drugs for required patient therapy. The Spectrum is used in conjunction with legally marked intravenous administration sets and medications provided by the user. The Master Drug Library is a software package that will add additional features to the Spectrum Infusion Pump. The Master Drug Library will permit electronic communications with the Spectrum pump and other external peripheral devices. The intended use of the Spectrum pump includes common drug error prevention, through the stand alone settings features of the pump. This includes drug parameter limits and associated drug name identification. With the Master Drug Library, the intended use is to reduce user errors associated with drug selection, drug dose rates, drug dose concentrations, and patient identification associated with the prescribed drug. recalled?
Yes. FDA published recall Z-1490-2012 on June 20, 2012. Reported hazard: Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.
What should I do if I have this product?
Voluntary: Firm initiated · Letter