FDAFDA · Food and drugs·Recall number Z-1492-2013

Stryker Instruments0400-750-000 T4 Toga, Pullover, Regular Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

Published June 12, 2013 · Recall initiated: May 23, 2013

Summary

FDA records medical device Class II recall Z-1492-2013, published June 12, 2013 and initiated May 23, 2013. Recalling firm: Stryker Instruments Div. of Stryker Corporation; Status: Terminated; Reported quantity: 3,000 units (300 boxes); Lot: 1005111, 1011282,1107491, 1107492. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Instruments Div. of Stryker Corporation
Distributor
USA Nationwide Distribution
Lot
1005111, 1011282,1107491, 1107492
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Stryker Instruments0400-750-000 T4 Toga, Pullover, Regular Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material. recalled?

Yes. FDA published recall Z-1492-2013 on June 12, 2013. Reported hazard: Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit