Stryker Instruments 0400-760-000 T4 Toga, Pullover, Large Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.
Published June 12, 2013 · Recall initiated: May 23, 2013
Summary
FDA records medical device Class II recall Z-1493-2013, published June 12, 2013 and initiated May 23, 2013. Recalling firm: Stryker Instruments Div. of Stryker Corporation; Status: Terminated; Reported quantity: 34,640 units (3,464 boxes); Lot: 1006134, 1012304, 1104308, 1007152, 1101020, 1105365, 1007170, 1102080, 1106421, 1008188, 1102147, 1106469, 1009205, 1103193, 1108628, 1010229, 1103198, 1108697, 1010253, 1102149, 1109789, 1011277, 1104248. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Instruments Div. of Stryker Corporation
- Distributor
- USA Nationwide Distribution
- Lot
- 1006134, 1012304, 1104308, 1007152, 1101020, 1105365, 1007170, 1102080, 1106421, 1008188, 1102147, 1106469, 1009205, 1103193, 1108628, 1010229, 1103198, 1108697, 1010253, 1102149, 1109789, 1011277, 1104248
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting,July 27, 2022
- FDANeptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is anMay 6, 2020
- FDAColor Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 1PRT(Single Port) QuickDecember 11, 2019
- FDAC2 Nerve Monitor, 8 Channel, 5140-508-280, UDI 07613327277319 Product Usage: The C2 Nerve Monitor System isOctober 9, 2019
- FDAStryker Instruments0400-750-000 T4 Toga, Pullover, Regular Sterile, EO; single use; Product Usage: ProtectionJune 12, 2013
- FDAStryker Instruments 0400-770-000 T4 Toga, Pullover, Ex-Large Sterile, EO; single use; Product Usage:June 12, 2013
Questions about this recall
Was Stryker Instruments 0400-760-000 T4 Toga, Pullover, Large Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material. recalled?
Yes. FDA published recall Z-1493-2013 on June 12, 2013. Reported hazard: Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit