FDAFDA · Food and drugs·Recall number Z-1497-2017

7F Prelude¿ Short Sheath Introducer. Catalog Numbers: K15-00070, K15-00170, PSS-7F-4-035MT, PSS-7F-4MT. Produc Usage: The Merit Prelude¿ Short Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis. The device is not indicated for long term vascular or hemodialysis access.

Published April 5, 2017 · Recall initiated: February 16, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Merit Medical Systems, Inc. announces a voluntary field action for sheaths included with the 7F Prelude¿ Short Sheath Introducers because the sheath marker tips may detach during use.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Merit Medical Systems, Inc.
Distributor
Nationwide and other foreign consignees. No Canadian or VA/govt/military consignees.
Lot
Lot Numbers: H1041469, H1041473, H1036880, H1041464.
Risk classification
Class I
Area (FDA)
device
Country
United States