Back to search
FDAFDA · Food and drugs·Recall number Z-1504-2024

Trilogy Evo, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.

Published April 24, 2024 · Recall initiated: March 19, 2024

Summary

FDA records medical device Class I recall Z-1504-2024, published April 24, 2024 and initiated March 19, 2024. Recalling firm: Philips Respironics, Inc.; Status: Ongoing; Reported quantity: 76,713 units; Lot: Model No. AU2110X15B, BL2110X15B, BL2200X15B, BR2110X18B, BR2200X18B, CA2110X12B, CA2200X12B, DE2100X13B, DE2110X13B, DE2200X13B, DS2000X11B, DS2100X11B, DS2110X11B, DS2200X11B, EE2100X15B, EE2110X15B, EE2200X15B, ES2100X15B, ES2110X15B, ES2200X15B, EU2100X15B, EU2100X19, EU2110X15B, EU2110X19, EU2200X15B, FR2100X14B, FR2110X14B, FR2200X14B, FX2100X15B, FX2200X15B, GB2110X15B, GB2200X15B, IA2100X15B, IA2110X15B, IN2100X15B, IN2100X19, IN2110X15B, IN2200X15B, IT2100X21B, IT2110X21B, IT2200X21B, JP2100X16B, JP2110X16B, KR2110X15B, KR2200X15B, LA2100X15B, LA2110X15B, LD2110X23B, ND2100X15B, ND2110X15B, ND2200X15B, RDS2110X11B, SP2100X26B, TR2110X15B, TR2200X15B, UDS2110X11B, VT2110X24B; UDI-DI: 606959051980, 606959055520, 606959055476, 606959052307, 606959054110, 606959051942, 606959055544, 606959054080, 606959054981, 606959055537, 606959055506, 606959054127, 606959055490, 606959051959, 606959055421, 606959051973, 606959055483, 606959055452, 606959055513, 606959060517, 606959055551, 606959060500; All units except those with software version 1.05.06.00. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Respironics, Inc.
Distributor
Domestic: Nationwide Distribution; Foreign: AE, AR, AT, AU, BD, BE, BG, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, US, UY, VN, YT, ZA, ZW.
Lot
Model No. AU2110X15B, BL2110X15B, BL2200X15B, BR2110X18B, BR2200X18B, CA2110X12B, CA2200X12B, DE2100X13B, DE2110X13B, DE2200X13B, DS2000X11B, DS2100X11B, DS2110X11B, DS2200X11B, EE2100X15B, EE2110X15B, EE2200X15B, ES2100X15B, ES2110X15B, ES2200X15B, EU2100X15B, EU2100X19, EU2110X15B, EU2110X19, EU2200X15B, FR2100X14B, FR2110X14B, FR2200X14B, FX2100X15B, FX2200X15B, GB2110X15B, GB2200X15B, IA2100X15B, IA2110X15B, IN2100X15B, IN2100X19, IN2110X15B, IN2200X15B, IT2100X21B, IT2110X21B, IT2200X21B, JP2100X16B, JP2110X16B, KR2110X15B, KR2200X15B, LA2100X15B, LA2110X15B, LD2110X23B, ND2100X15B, ND2110X15B, ND2200X15B, RDS2110X11B, SP2100X26B, TR2110X15B, TR2200X15B, UDS2110X11B, VT2110X24B; UDI-DI: 606959051980, 606959055520, 606959055476, 606959052307, 606959054110, 606959051942, 606959055544, 606959054080, 606959054981, 606959055537, 606959055506, 606959054127, 606959055490, 606959051959, 606959055421, 606959051973, 606959055483, 606959055452, 606959055513, 606959060517, 606959055551, 606959060500; All units except those with software version 1.05.06.00.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Trilogy Evo, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. recalled?

Yes. FDA published recall Z-1504-2024 on April 24, 2024. Reported hazard: Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

What should I do if I have this product?

Voluntary: Firm initiated · Letter