Back to search
FDAFDA · Food and drugs·Recall number Z-1505-2015

Specialty MiniScope Mech. Set Size 18, Specialty MiniScope Mech. Set Size 21, and Specialty MiniScope Mech. Set Size 24 A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.

Published April 29, 2015 · Recall initiated: March 9, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Specialty Appliance Works, Inc.
Distributor
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
Lot
Product No. 90870, 90876 and 90879
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.