Back to search
FDAFDA · Food and drugs·Recall number Z-1511-2015

TS ROC Crown Component 4UR (5/pk), TS ROC Crown Component 5UR (5/pk), TS ROC Crown Component 6UR (5/pk), TS ROC Crown Component 7UR (5/pk), TS ROC Crown Component 8UR (5/pk), TS ROC Crown Component 4LR (5/pk), TS ROC Crown Component 5LR (5/pk), TS ROC Crown Component 6LR (5/pk), TS ROC Crown Component 7LR (5/pk), TS ROC Crown Component 8LR (5/pk), TS ROC Crown Component 4UL (5/pk), TS ROC Crown Component 5UL (5/pk), TS ROC Crown Component 6UL (5/pk), TS ROC Crown Component 7UL (5/pk), TS ROC Crown Component 8UL (5/pk), TS ROC Crown Component 4LL (5/pk), TS ROC Crown Component 5LL (5/pk), TS ROC Crown Component 6LL (5/pk), TS ROC Crown Component 7LL (5/pk), TS ROC Crown Component 8LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.

Published April 29, 2015 · Recall initiated: March 9, 2015

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Specialty Appliance Works, Inc.
Distributor
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
Lot
Product No. 91720, 91721, 91722, 91723, 91724, 91725, 91726, 91727, 91728, 91729, 91730, 91731, 91732, 91733, 91734, 91735, 91736, 91737, 91738, and 91739
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.