TS ROC Crown Component 4UR (5/pk), TS ROC Crown Component 5UR (5/pk), TS ROC Crown Component 6UR (5/pk), TS ROC Crown Component 7UR (5/pk), TS ROC Crown Component 8UR (5/pk), TS ROC Crown Component 4LR (5/pk), TS ROC Crown Component 5LR (5/pk), TS ROC Crown Component 6LR (5/pk), TS ROC Crown Component 7LR (5/pk), TS ROC Crown Component 8LR (5/pk), TS ROC Crown Component 4UL (5/pk), TS ROC Crown Component 5UL (5/pk), TS ROC Crown Component 6UL (5/pk), TS ROC Crown Component 7UL (5/pk), TS ROC Crown Component 8UL (5/pk), TS ROC Crown Component 4LL (5/pk), TS ROC Crown Component 5LL (5/pk), TS ROC Crown Component 6LL (5/pk), TS ROC Crown Component 7LL (5/pk), TS ROC Crown Component 8LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
Published April 29, 2015 · Recall initiated: March 9, 2015
Summary
FDA records medical device Class II recall Z-1511-2015, published April 29, 2015 and initiated March 9, 2015. Recalling firm: Specialty Appliance Works, Inc.; Status: Terminated; Reported quantity: 14,428 Universal Nuts; Lot: Product No. 91720, 91721, 91722, 91723, 91724, 91725, 91726, 91727, 91728, 91729, 91730, 91731, 91732, 91733, 91734, 91735, 91736, 91737, 91738, and 91739. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Specialty Appliance Works, Inc.
- Distributor
- Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
- Lot
- Product No. 91720, 91721, 91722, 91723, 91724, 91725, 91726, 91727, 91728, 91729, 91730, 91731, 91732, 91733, 91734, 91735, 91736, 91737, 91738, and 91739
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFluoxetine Capsules USP, 20 mg, packaged in a) 500-count bottles (NDC 50111-648-02), and (b) 1000- countMay 6, 2015
- FDASpecialty MiniScope Mech. Set Size 18, Specialty MiniScope Mech. Set Size 21, and Specialty MiniScope Mech.April 29, 2015
- FDATS Herbst Nuts with Backing (10/pk) and TS Herbst Nuts Universal (10/pk) A Herbst Appliance is used to affectApril 29, 2015
- FDAOptiMedica Catalys Precision Laser System; Model No(s). Catalys-U (US), and Catalys-I (Outside US). ProductApril 29, 2015
- FDAThe PCA 3 infusion pump allows clinicians, or patients, to (self) Administer analgesia safely and effectivelyApril 29, 2015
- FDASentec, V-Sign Sensor 2 (REF VS-A/P/N), Reusable / Nonsterile / Waterproof, SenTec AG, CH-4106, Therwil,April 29, 2015
Questions about this recall
Was TS ROC Crown Component 4UR (5/pk), TS ROC Crown Component 5UR (5/pk), TS ROC Crown Component 6UR (5/pk), TS ROC Crown Component 7UR (5/pk), TS ROC Crown Component 8UR (5/pk), TS ROC Crown Component 4LR (5/pk), TS ROC Crown Component 5LR (5/pk), TS ROC Crown Component 6LR (5/pk), TS ROC Crown Component 7LR (5/pk), TS ROC Crown Component 8LR (5/pk), TS ROC Crown Component 4UL (5/pk), TS ROC Crown Component 5UL (5/pk), TS ROC Crown Component 6UL (5/pk), TS ROC Crown Component 7UL (5/pk), TS ROC Crown Component 8UL (5/pk), TS ROC Crown Component 4LL (5/pk), TS ROC Crown Component 5LL (5/pk), TS ROC Crown Component 6LL (5/pk), TS ROC Crown Component 7LL (5/pk), TS ROC Crown Component 8LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance. recalled?
Yes. FDA published recall Z-1511-2015 on April 29, 2015. Reported hazard: The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.
What should I do if I have this product?
Voluntary: Firm initiated · Letter