Giraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.
Published March 29, 2017 · Recall initiated: November 18, 2016
Summary
FDA records medical device Class II recall Z-1506-2017, published March 29, 2017 and initiated November 18, 2016. Recalling firm: Ohmeda Medical; Status: Terminated; Reported quantity: 5,993 (5,080 US; 913 OUS); Lot: See Attached Spreadsheat. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ohmeda Medical
- Distributor
- Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
- Lot
- See Attached Spreadsheat
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGiraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant incubator. IncubatorsMarch 29, 2017
- FDACarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled thermal environment forMarch 29, 2017
- FDAGiraffe Bedded Warmer, Panda Warmer Modification to Giraffe and Panda Warmers (Giraffe and PandaMarch 29, 2017
- FDAGiraffe Bedded Warmer, Panda Warmer Modifiction to Giraffe and Panda Warmer- Infant radiant warmers provideMarch 29, 2017
- FDAGiraffe OmniBed Ohmeda Medical Giraffe OmniBed- The OmniBed is a combination of an infant incubator and anMarch 29, 2017
- FDAGiraffe Spot PT Ohmeda Medical Spot PT Lite Phototherapy System- The Spot PT Lite Phototherapy System providesMarch 29, 2017
Questions about this recall
Was Giraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant. recalled?
Yes. FDA published recall Z-1506-2017 on March 29, 2017. Reported hazard: Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail