CarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most incubators can be used in two operating modes: 1. Air Control: The clinician sets the appropriate air temperature for maintaining the desired patient temperature. The air temperature is initially selected based on the clinician s training and experience and then is adjusted based on the patient s needs and clinical status. 2. Patient control: The clinician sets the desired patient temperature. A skin temperature probe sense the patient temperature and feeds this information to the controller of the incubator. Thecontroller then adjusts the heater output to maintain the patient temperature at the set value. These adjustments to the heater output are made in such a way to gradually change the patient s temperature while minimizing overshooting and patient stress. Incubators have alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient s temperature from the set value. Incubators may incorporate other features, such as humidification of the infant environment, tilting of the bed, oxygen supply, and data output to remote monitors or nurse call systems
Published March 29, 2017 · Recall initiated: November 18, 2016
Summary
FDA records medical device Class II recall Z-1510-2017, published March 29, 2017 and initiated November 18, 2016. Recalling firm: Ohmeda Medical; Status: Terminated; Reported quantity: 5,993 (5,080 US; 913 OUS); Lot: See Attached Spreadsheat. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ohmeda Medical
- Distributor
- Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
- Lot
- See Attached Spreadsheat
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGiraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat inMarch 29, 2017
- FDAGiraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant incubator. IncubatorsMarch 29, 2017
- FDAGiraffe Bedded Warmer, Panda Warmer Modification to Giraffe and Panda Warmers (Giraffe and PandaMarch 29, 2017
- FDAGiraffe Bedded Warmer, Panda Warmer Modifiction to Giraffe and Panda Warmer- Infant radiant warmers provideMarch 29, 2017
- FDAMammotome Revolve Dual Vacuum Assist Biopsy System The device is used to sample breast tissue that has beenMarch 29, 2017
- FDAPoly Component Trial, UCCS, packaged in the following sizes and configurations: a) Poly Comp Trial - UCCSMarch 29, 2017
Questions about this recall
Was CarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most incubators can be used in two operating modes: 1. Air Control: The clinician sets the appropriate air temperature for maintaining the desired patient temperature. The air temperature is initially selected based on the clinician s training and experience and then is adjusted based on the patient s needs and clinical status. 2. Patient control: The clinician sets the desired patient temperature. A skin temperature probe sense the patient temperature and feeds this information to the controller of the incubator. Thecontroller then adjusts the heater output to maintain the patient temperature at the set value. These adjustments to the heater output are made in such a way to gradually change the patient s temperature while minimizing overshooting and patient stress. Incubators have alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient s temperature from the set value. Incubators may incorporate other features, such as humidification of the infant environment, tilting of the bed, oxygen supply, and data output to remote monitors or nurse call systems recalled?
Yes. FDA published recall Z-1510-2017 on March 29, 2017. Reported hazard: Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail