Back to search
FDAFDA · Food and drugs·Recall number Z-1517-2020

Custom procedure trays containing medical components needed to facilitate patient procedures and to perform minor and major patient surgeries. Multiple item and lot numbers.

Published March 25, 2020 · Recall initiated: February 18, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Avid Medical, Inc.
Distributor
US: AZ, CA, IL, IN, MD, MS, NJ, NV, NY, TN, TX WI, VA
Lot
Item Codes: ADBR043-08, ADCO001-20, ADCO001-20, ADCO002-15, ADCO005-29, ADCO007-21, ADVC015-23, ADVC032-21, ADVC046-12, ADVH073-13, ADVM040-07, ADVM050-02, SHER020-10, SINA223, VAHT026-07, YAVA115-04, YAVA130, YAVA131-03. Lot numbers: 1333107, 1333471, 1336767, 1337370, 1339098, 1339331, 1342016, 1347760, 1347896, 1348407, 1353068, 1353419, 1354811, 1357043, 1359839, 1360420, 1361445, 1363245, 1365971, 1367018, 1367137, 1367218, 1367987, 1368726, 1369130, 1370525, 1370700, 1370929, 1371222, 1372022, 1372310, 1372407, 1373347, 1374864, 1375307, 1375787, 1376196, 1376397, 1378227, 1380392, 1380874, 1381757, 1382980, 1383533, 1384526, 1385192. Expiration Dates: 9/30/2020-10/17/2022
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card