Custom procedure trays containing medical components needed to facilitate patient procedures and to perform minor and major patient surgeries. Multiple item and lot numbers.

Published March 25, 2020 · Recall initiated: February 18, 2020

Summary

FDA records medical device Class II recall Z-1518-2020, published March 25, 2020 and initiated February 18, 2020. Recalling firm: Avid Medical, Inc.; Status: Terminated; Reported quantity: 400 each; Lot: Item Codes: ADVC065-18, ADVC068-21, ADVH078-17, ADVM042-07. Lot Numbers: 1324410, 1344158, 1356561, 1366768, 1372012, 1373518, 1375096, 1377349, 1378072, 1378073, 1379861, 1381271, 1381272. Expiration dates: 05/01/2020-09/30/2021. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Avid Medical, Inc.
Distributor
US: AZ, CA, IL, IN, MD, MS, NJ, NV, NY, TN, TX WI, VA
Lot
Item Codes: ADVC065-18, ADVC068-21, ADVH078-17, ADVM042-07. Lot Numbers: 1324410, 1344158, 1356561, 1366768, 1372012, 1373518, 1375096, 1377349, 1378072, 1378073, 1379861, 1381271, 1381272. Expiration dates: 05/01/2020-09/30/2021.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Custom procedure trays containing medical components needed to facilitate patient procedures and to perform minor and major patient surgeries. Multiple item and lot numbers. recalled?

Yes. FDA published recall Z-1518-2020 on March 25, 2020. Reported hazard: The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.