TX1 Tissue Removal System disposable handpieces. Product Usage: The TX1 Tissue Removal System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery.
Published June 19, 2013 · Recall initiated: April 30, 2013
Summary
FDA records medical device Class II recall Z-1518-2013, published June 19, 2013 and initiated April 30, 2013. Recalling firm: American Optisurgical Inc; Status: Terminated; Reported quantity: 1,436 units; Lot: Lot Numbers: 00312-07, 00713-06, 00912-01, 01813-02, 02413-04, 03113-02, 04213-01, 05113-01, 05113-02, 05512-03, 05713-03, 06313-05, 06512-08, 06513-02, 07213-04, 07912-06, 08212-07, 09412-03, 11412-06, 11612-05, 12912-04, 14512-07, 15312-03, 16412-07, 17412-01, 18712-06, 19812-13, 20212-03, 21412-03, 22212-03, 23412-06, 24112-06, 24212-04, 24912-13, 25712-01, 26912-01, 27712-05, 28512-03, 28912-03, 29612-01, 30512-02, 31112-05, 32112-03, 33112-01, 34712-05, 36612-01. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
American Optisurgical, Inc. is recalling certain lots of TX1 Tissue Removal System disposable handpiece due to inventory being released into distribution without a cleared FDA 510(k) to cover recent product changes. Use of this product may lead to an underperformance of the device.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Optisurgical Inc
- Distributor
- Worldwide Disitribution - USA Nationwide in the states of Alabama, Arkansas, Arizona, California, Colorado, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming and the countries of Quezon City, Philippines, Singapore, and Panama City, Panama.
- Lot
- Lot Numbers: 00312-07, 00713-06, 00912-01, 01813-02, 02413-04, 03113-02, 04213-01, 05113-01, 05113-02, 05512-03, 05713-03, 06313-05, 06512-08, 06513-02, 07213-04, 07912-06, 08212-07, 09412-03, 11412-06, 11612-05, 12912-04, 14512-07, 15312-03, 16412-07, 17412-01, 18712-06, 19812-13, 20212-03, 21412-03, 22212-03, 23412-06, 24112-06, 24212-04, 24912-13, 25712-01, 26912-01, 27712-05, 28512-03, 28912-03, 29612-01, 30512-02, 31112-05, 32112-03, 33112-01, 34712-05, 36612-01
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: AFX" Introducer System, Model Number S17-45. Lot Numbers: 1079840, 1079843, 1079844, 1079845.June 19, 2013
- FDADePuy Glenosphere orientation guide Product Usage: The glenosphere orientation guide is an instrument used inJune 19, 2013
- FDAJumbo Barcelona in-shell hazelnuts and Large Barcelona in-shell hazelnuts, raw, Hazelnut Growers of OregonJune 19, 2013
- FDAFAST Procedure Kit. Intended for use as a Minor Surgical Procedures Tray.June 12, 2013
- FDAAmerican Optisurgical's TX1 Tissue Removal System Console. The console is a software unit that is indicatedJune 12, 2013
- FDATX1 Tissue Removal System Console. Model: TX1 TX1 Tissue Removal System is an ultrasonic aspirator thatJune 27, 2012
Questions about this recall
Was TX1 Tissue Removal System disposable handpieces. Product Usage: The TX1 Tissue Removal System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery. recalled?
Yes. FDA published recall Z-1518-2013 on June 19, 2013. Reported hazard: American Optisurgical, Inc. is recalling certain lots of TX1 Tissue Removal System disposable handpiece due to inventory being released into distribution without a cleared FDA 510(k) to cover recent product changes. Use of this product may lead to an underperformance of the device.
What should I do if I have this product?
Voluntary: Firm initiated · Letter