TX1 Tissue Removal System Console. Model: TX1 TX1 Tissue Removal System is an ultrasonic aspirator that emulsifies and removes soft tissue. The TX1 Console provides control over each mode of operation, as well as the user interface via LCD touchscreen and audible tones for confirmation of user selections. Indiciated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery.
Published June 27, 2012 · Recall initiated: February 22, 2012
Summary
FDA records medical device Class II recall Z-1817-2012, published June 27, 2012 and initiated February 22, 2012. Recalling firm: American Optisurgical Inc; Status: Terminated; Reported quantity: 5units; Lot: Product Codes: LFL, GEI, Serial #: 55400001, 55400002, 55400003. Part Order No: 554-0034-000 510(K) Mp.: K101561. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Two reasons for recall. 1. Incorrect pressure relief valve installed inside console, leading to a remote chance of a risk to health. 2. Technical Specifications, under Irrigation, the Operating Pressure value is incorrect.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Optisurgical Inc
- Distributor
- Nationwide distribution: MN only.
- Lot
- Product Codes: LFL, GEI, Serial #: 55400001, 55400002, 55400003. Part Order No: 554-0034-000 510(K) Mp.: K101561
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATX1 Tissue Removal System disposable handpieces. Product Usage: The TX1 Tissue Removal System is indicated forJune 19, 2013
- FDAFAST Procedure Kit. Intended for use as a Minor Surgical Procedures Tray.June 12, 2013
- FDAAmerican Optisurgical's TX1 Tissue Removal System Console. The console is a software unit that is indicatedJune 12, 2013
- FDAStryker Howmedica Osteonics Accolade Offset Rasp Handle Howmedica Osteonics Corp. 325 Corporate Drive, Mahwah,June 27, 2012
- FDAPhilips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon StationaryJune 27, 2012
- FDAPrepStain GN 480 Disposable Kit Catalog/Model 490542 This is a complete kit, per the firm's explanation, thatJune 27, 2012
Questions about this recall
Was TX1 Tissue Removal System Console. Model: TX1 TX1 Tissue Removal System is an ultrasonic aspirator that emulsifies and removes soft tissue. The TX1 Console provides control over each mode of operation, as well as the user interface via LCD touchscreen and audible tones for confirmation of user selections. Indiciated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery. recalled?
Yes. FDA published recall Z-1817-2012 on June 27, 2012. Reported hazard: Two reasons for recall. 1. Incorrect pressure relief valve installed inside console, leading to a remote chance of a risk to health. 2. Technical Specifications, under Irrigation, the Operating Pressure value is incorrect.
What should I do if I have this product?
Voluntary: Firm initiated · Letter