Back to search
FDAFDA · Food and drugs·Recall number Z-1518-2024

Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098

Published April 17, 2024 · Recall initiated: March 13, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Intuitive Surgical, Inc.
Distributor
US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea
Lot
Part/UDI-DI: 380601/00886874114605. SP1098 Serial Numbers: SP0202, SP0253, SP0254, SP0219, SP0228, SP0231, SP0238, SP0273, SP0208, SP0226, SP0240, SP0247, SP0249, SP0256, SP0257, SP0268, SP0270, SP0272, SP0221, SP0223, SP0225, SP0229, SP0230, SP0232, SP0233, SP0235, SP0236, SP0237, SP0239, SP0241, SP0242, SP0243, SP0244, SP0245, SP0248, SP0250, SP0251, SP0252, SP0255, SP0258, SP0259, SP0260, SP0261, SP0262, SP0263, SP0264, SP0265, SP0266, SP0269, SP0274, SP0278, SP0280, SP0282
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card