Endo-Model Replacement Plateau; Item Number: 15-0027/11;
Published March 18, 2026 · Recall initiated: January 12, 2026
Summary
FDA records medical device Class II recall Z-1521-2026, published March 18, 2026 and initiated January 12, 2026. Recalling firm: Waldemar Link GmbH & Co. KG (Mfg Site); Status: Ongoing; Reported quantity: 2 units (OUS only); Lot: Item Number: 15-0027/11; UDI-DI: 04026575431953; Serial/Lot Number: 2549126, 2549127. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Distributor
- Worldwide - US Nationwide distribution in the state of NJ and the countries of Belarus, Canada, Switzerland , Colombia, Germany, Spain, United Kingdom, Hungary, India, Italy, Netherlands, Romania, Saudi Arabia.
- Lot
- Item Number: 15-0027/11; UDI-DI: 04026575431953; Serial/Lot Number: 2549126, 2549127;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAEndo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item NumberAugust 19, 2026
- FDAPenner Pacific Bathing Spa, Model Numbers 360020-1PMarch 18, 2026
- FDAGEM Premier 5000 PAK, Part No. 00055360004.March 18, 2026
- FDAEndo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component,July 24, 2024
- FDAEndo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02,July 24, 2024
- FDASPII Model Lubinus, Long Stem Prosthesis Standard Neck, cementedMarch 20, 2024
Questions about this recall
Was Endo-Model Replacement Plateau; Item Number: 15-0027/11; recalled?
Yes. FDA published recall Z-1521-2026 on March 18, 2026. Reported hazard: The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
What should I do if I have this product?
Voluntary: Firm initiated · Letter