Back to search
FDAFDA · Food and drugs·Recall number Z-2934-2026

Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16

Published August 19, 2026 · Recall initiated: July 6, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Waldemar Link GmbH & Co. KG (Mfg Site)
Distributor
Domestic US distribution to Indiana. OUS distribution to Belarus and Germany.
Lot
UDI-DI 04026575432134 Lot for Implant REF 15-0027/16: 2604069
Risk classification
Class II
Area (FDA)
device
Country
Germany