Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16

Published August 19, 2026 · Recall initiated: July 6, 2026

Summary

FDA records medical device Class II recall Z-2934-2026, published August 19, 2026 and initiated July 6, 2026. Recalling firm: Waldemar Link GmbH & Co. KG (Mfg Site); Status: Ongoing; Reported quantity: 1 US; 9 OUS; Lot: UDI-DI 04026575432134 Lot for Implant REF 15-0027/16: 2604069. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Waldemar Link GmbH & Co. KG (Mfg Site)
Distributor
Domestic US distribution to Indiana. OUS distribution to Belarus and Germany.
Lot
UDI-DI 04026575432134 Lot for Implant REF 15-0027/16: 2604069
Risk classification
Class II
Area (FDA)
device
Country
Germany

Related recalls

Questions about this recall

Was Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16 recalled?

Yes. FDA published recall Z-2934-2026 on August 19, 2026. Reported hazard: The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.

What should I do if I have this product?

Voluntary: Firm initiated · Letter