Back to search
FDAFDA · Food and drugs·Recall number Z-1527-2013

BARD Biopsy Systems (A Business Unit of Bard Peripheral Vascular, Inc.), UltraClip¿ Dual Trigger Breast Tissue Marker 17g X 10cm needle, Ultrasound Enhanced Ribbon Marker 5 Pack, Sterile, Single Use, Non-pyrogenic, Rx only; Bard Peripheral Vascular, Inc., 1625 West 3rd Street, Tempe, AZ 85281 USA. PK4313800 Rev. 1 01/11 The UltraClip Dual Trigger breast tissue marker is a sterile, single use device comprised of a disposable introducer, and a metal implantable tissue marker with polyvinyl alcohol (PVA).

Published June 19, 2013 · Recall initiated: January 23, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard¿ UltraClip¿ Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization process (non sterile).

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Bard Peripheral Vascular Inc
Distributor
Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KY, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, TX, VA, WA, WV and WY.
Lot
PC 863017D; Lot: HUWL0010
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card