VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102).
Published June 19, 2013 · Recall initiated: May 24, 2013
Summary
FDA records medical device Class II recall Z-1531-2013, published June 19, 2013 and initiated May 24, 2013. Recalling firm: International Biomedical, Ltd.; Status: Terminated; Reported quantity: 7,000; Lot: Model/Part Number: 209-2113, or as part of kits 777-3103 and 777-3102. Collection Bag Lot numbers: 4389, 4465, 4478, 4516, 4530, 4430, 4489. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
International Biomedical has received reports of interference between the collection bag and the extension set connectors. Evaluation of the affected components indicates that the collection bag connector can become occluded due to an over-tightening of the luer lock by the end user. To alleviate the occurrence of this issue in the fixture, the connector on the collection bag has been improved to
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- International Biomedical, Ltd.
- Distributor
- Nationwide Distribution.
- Lot
- Model/Part Number: 209-2113, or as part of kits 777-3103 and 777-3102. Collection Bag Lot numbers: 4389, 4465, 4478, 4516, 4530, 4430, 4489
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack with Tyvek lid. The ECHELON andJune 19, 2013
- FDAembla Embletta Gold (It functions as a Polysomnography Recorder and Ventilator Effort Recorder). The EmblettaJune 19, 2013
- FDAUSHER® Support Catheter- Angled Tip WL130cm, RI 7F; REF USH07AT; PK1022-01A; Rx only, Non-Pyrogenic, Sterile;June 19, 2013
- FDABARD PERIPHERAL VASCULAR VACCESS® PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess®June 19, 2013
- FDAMICROSHEATH® - 0.066 (1.7mm) proximal, 0.064 (1.6mm); WL 123cm, RGC 8F Guide; minimum ID: 0.086"/June 19, 2013
- FDA90 Degree Cannulated Infant Blade Plate 25mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics LockingJune 19, 2013
Questions about this recall
Was VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102). recalled?
Yes. FDA published recall Z-1531-2013 on June 19, 2013. Reported hazard: International Biomedical has received reports of interference between the collection bag and the extension set connectors. Evaluation of the affected components indicates that the collection bag connector can become occluded due to an over-tightening of the luer lock by the end user. To alleviate the occurrence of this issue in the fixture, the connector on the collection bag has been improved to
What should I do if I have this product?
Voluntary: Firm initiated · Letter