Back to search
FDAFDA · Food and drugs·Recall number Z-1531-2013

VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102).

Published June 19, 2013 · Recall initiated: May 24, 2013

Summary

FDA records medical device Class II recall Z-1531-2013, published June 19, 2013 and initiated May 24, 2013. Recalling firm: International Biomedical, Ltd.; Status: Terminated; Reported quantity: 7,000; Lot: Model/Part Number: 209-2113, or as part of kits 777-3103 and 777-3102. Collection Bag Lot numbers: 4389, 4465, 4478, 4516, 4530, 4430, 4489. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

International Biomedical has received reports of interference between the collection bag and the extension set connectors. Evaluation of the affected components indicates that the collection bag connector can become occluded due to an over-tightening of the luer lock by the end user. To alleviate the occurrence of this issue in the fixture, the connector on the collection bag has been improved to

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
International Biomedical, Ltd.
Distributor
Nationwide Distribution.
Lot
Model/Part Number: 209-2113, or as part of kits 777-3103 and 777-3102. Collection Bag Lot numbers: 4389, 4465, 4478, 4516, 4530, 4430, 4489
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile LVM blood gas sensor with the LVM Monitor. The collection bag is packaged in a tyvek pouch which is included in either replacement kit (777-3103) or LVM sensor kit (777-3102). recalled?

Yes. FDA published recall Z-1531-2013 on June 19, 2013. Reported hazard: International Biomedical has received reports of interference between the collection bag and the extension set connectors. Evaluation of the affected components indicates that the collection bag connector can become occluded due to an over-tightening of the luer lock by the end user. To alleviate the occurrence of this issue in the fixture, the connector on the collection bag has been improved to

What should I do if I have this product?

Voluntary: Firm initiated · Letter