Back to search
FDAFDA · Food and drugs·Recall number Z-1534-2013

90 Degree Cannulated Infant Blade Plate 25mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, intertrochanteric valgus osteotomies, proximal and distal tibial osteotomies and humeral fractures and osteotomies.

Published June 19, 2013 · Recall initiated: April 19, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
OrthoPediatrics Corp
Distributor
USA Nationwide Distribution
Lot
All lot numbers are affected. Catalog No. 00-1200-1000
Risk classification
Class II
Area (FDA)
device
Country
United States