FDAFDA · Food and drugs·Recall number Z-1543-2024

Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716

Published April 24, 2024 · Recall initiated: December 19, 2023

Summary

FDA records medical device Class II recall Z-1543-2024, published April 24, 2024 and initiated December 19, 2023. Recalling firm: Smiths Medical ASD Inc.; Status: Ongoing; Reported quantity: 1842 guide barrel clamps; Lot: Lot Numbers: P0407365, P0486670, P0561740. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
Worldwide distribution.
Lot
Lot Numbers: P0407365, P0486670, P0561740
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716 recalled?

Yes. FDA published recall Z-1543-2024 on April 24, 2024. Reported hazard: Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

What should I do if I have this product?

Voluntary: Firm initiated · Letter