FDAFDA · Food and drugs·Recall number Z-1544-2014
12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.
Published May 14, 2014 · Recall initiated: April 16, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Stryker Endoscopy
- Distributor
- Worldwide distribution
- Lot
- Model number: 0242000012; Lot/serial numbers: 021464-2, 39824-1-1, 4014, 643893-1-4, 27419, 42901-1-1, 35728-2-1, 48639-1-8, 021464-3, 40288-1-1, 43020, 43893-1-6, 27701, 43104-1-1, 36175-1-1, 48956-1-1, 23027, 40629-3-,1, 43622, 43893-1-7, 29129, 43270-1-1, 36421-1-1, 48956-1-1A, 23398, 40721-1-1, 50476, 43893-1-8, 29670, 43523-1-1, 36424-1-1, 51304-1-1, 23542, 41231-4-1, 54990, 43893-1-9, 30250, 43523-1-1A, 37251-1-1, 51304-1-1A, 23846, 41478-1-1, 7J1, 48639-1-2, 30919, 43674-1-1, 37392-1-1, 52123-1-1, 24073, 41745-1-1, 123, 48639-1-3, 3218., 43893-1-1, 37794-1-1, 52956-1-1, 24511, 42086-1-1, 28646-1-1, 48639-1-4, 33226, 43893-1-10, 38867-1-1, 54346-1-1, 26445, 42306-1-1, 29005-1-1, 48639-1-5, 33961, 43893-1-1, 1SS15327, 26843, 42526-1-1, 30384-1-1, 48639-1-7, 35933, 43893-1-12, 37991, 36610, 43893-1-13, 39219, 43893-1-2, 43893-1-14.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States