12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.
Published May 14, 2014 · Recall initiated: April 16, 2014
Summary
FDA records medical device Class II recall Z-1544-2014, published May 14, 2014 and initiated April 16, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 2591 trays; Lot: Model number: 0242000012; Lot/serial numbers: 021464-2, 39824-1-1, 4014, 643893-1-4, 27419, 42901-1-1, 35728-2-1, 48639-1-8, 021464-3, 40288-1-1, 43020, 43893-1-6, 27701, 43104-1-1, 36175-1-1, 48956-1-1, 23027, 40629-3-,1, 43622, 43893-1-7, 29129, 43270-1-1, 36421-1-1, 48956-1-1A, 23398, 40721-1-1, 50476, 43893-1-8, 29670, 43523-1-1, 36424-1-1, 51304-1-1, 23542, 41231-4-1, 54990, 43893-1-9, 30250, 43523-1-1A, 37251-1-1, 51304-1-1A, 23846, 41478-1-1, 7J1, 48639-1-2, 30919, 43674-1-1, 37392-1-1, 52123-1-1, 24073, 41745-1-1, 123, 48639-1-3, 3218., 43893-1-1, 37794-1-1, 52956-1-1, 24511, 42086-1-1, 28646-1-1, 48639-1-4, 33226, 43893-1-10, 38867-1-1, 54346-1-1, 26445, 42306-1-1, 29005-1-1, 48639-1-5, 33961, 43893-1-1, 1SS15327, 26843, 42526-1-1, 30384-1-1, 48639-1-7, 35933, 43893-1-12, 37991, 36610, 43893-1-13, 39219, 43893-1-2, 43893-1-14. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Endoscopy
- Distributor
- Worldwide distribution
- Lot
- Model number: 0242000012; Lot/serial numbers: 021464-2, 39824-1-1, 4014, 643893-1-4, 27419, 42901-1-1, 35728-2-1, 48639-1-8, 021464-3, 40288-1-1, 43020, 43893-1-6, 27701, 43104-1-1, 36175-1-1, 48956-1-1, 23027, 40629-3-,1, 43622, 43893-1-7, 29129, 43270-1-1, 36421-1-1, 48956-1-1A, 23398, 40721-1-1, 50476, 43893-1-8, 29670, 43523-1-1, 36424-1-1, 51304-1-1, 23542, 41231-4-1, 54990, 43893-1-9, 30250, 43523-1-1A, 37251-1-1, 51304-1-1A, 23846, 41478-1-1, 7J1, 48639-1-2, 30919, 43674-1-1, 37392-1-1, 52123-1-1, 24073, 41745-1-1, 123, 48639-1-3, 3218., 43893-1-1, 37794-1-1, 52956-1-1, 24511, 42086-1-1, 28646-1-1, 48639-1-4, 33226, 43893-1-10, 38867-1-1, 54346-1-1, 26445, 42306-1-1, 29005-1-1, 48639-1-5, 33961, 43893-1-1, 1SS15327, 26843, 42526-1-1, 30384-1-1, 48639-1-7, 35933, 43893-1-12, 37991, 36610, 43893-1-13, 39219, 43893-1-2, 43893-1-14.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,December 24, 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDARetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 801, 802, 803, 803A: 80 " bed with 1, 2, or 3,May 14, 2014
- FDACLINITEK Status+ Analyzer (portable urine chemistry analyzer). Catalog number 10490946.May 14, 2014
Questions about this recall
Was 12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process. recalled?
Yes. FDA published recall Z-1544-2014 on May 14, 2014. Reported hazard: Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter