RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 801, 802, 803, 803A: 80 " bed with 1, 2, or 3, motors and autocontour 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
Published May 14, 2014 · Recall initiated: April 18, 2014
Summary
FDA records medical device Class II recall Z-1543-2014, published May 14, 2014 and initiated April 18, 2014. Recalling firm: Med-Mizer, Inc.; Status: Terminated; Reported quantity: 2980; Lot: Model 801 : 50000-50018; Model 802n : 600000-600789 Model 803: 500000-500342; 803A 700000-701724. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
All configurations of the SS Retractabed, Clinical Contour, made prior to June 7, 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on t
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Med-Mizer, Inc.
- Distributor
- USA (nationwide)
- Lot
- Model 801 : 50000-50018; Model 802n : 600000-600789 Model 803: 500000-500342; 803A 700000-701724
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763May 14, 2014
- FDARetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 799, 120 V, ...Max Safe Working Load 450 LBS.May 14, 2014
- FDA12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgicalMay 14, 2014
- FDACLINITEK Status+ Analyzer (portable urine chemistry analyzer). Catalog number 10490946.May 14, 2014
- FDAAlere Triage BNP Test for the Beckman Coulter Access Family of Immunoassay Systems, Model Number 98200, LotMay 14, 2014
- FDAAlere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR MonitoringMay 14, 2014
Questions about this recall
Was RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 801, 802, 803, 803A: 80 " bed with 1, 2, or 3, motors and autocontour 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care. recalled?
Yes. FDA published recall Z-1543-2014 on May 14, 2014. Reported hazard: All configurations of the SS Retractabed, Clinical Contour, made prior to June 7, 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on t
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit