TX1 Tissue Removal System. TX1 Procedure Pack P/N 554-1003-001 (outer box), TX1 MicroTip P/N 554-1002-001 (inner tray). For use with the TX System Console as an ultrasonic surgical aspirator of soft tissue.
Published May 14, 2014 · Recall initiated: April 14, 2014
Summary
FDA records medical device Class II recall Z-1552-2014, published May 14, 2014 and initiated April 14, 2014. Recalling firm: Tenex Health Inc; Status: Terminated; Reported quantity: 9,927 units; Lot: Lot Numbers: 00614-03, 06513-02, 24913-01, 00713-06, 06514-02, 25313-01, 00814-02, 06914-01, 25513-02, 01114-02, 07213-04, 25613-01, 01514-03, 07713-18, 26113-06, 01714-04, 08113-04, 26313-04, 01813-02, 08413-13, 26713-04, 02314-08, 09313-05, 26813-01, 02413-04, 10213-07, 27013-10, 02714-02, 10613-01, 27613-02, 02914-01, 11213-07, 28013-05, 03014-02, 11613-06, 28313-02, 03113-02, 12113-02, 29513-03, 03114-02, 12713-03, 29713-01, 03414-03, 13413-01, 30913-07, 03514-05, 14413-05, 31113-08, 03714-03, 15013-04, 31513-02, 03814-06, 15413-02, 31713-05, 04114-02, 15413-03, 32213-01, 04213-01, 16513-04, 32413-01, 04214-05, 17013-02, 32613-01, 04414-03, 17013-09, 32913-07, 04514-02, 17613-08, 33613-09, 04814-07, 17913-16, 33813-05, 05014-02, 18413-05, 34013-03, 05113-01, 18413-06, 34313-01, 05113-02, 18613-03, 34413-04, 05114-09, 19213-03, 34513-04, 05514-01, 19713-04, 34713-08, 05614-04, 20713-06, 35113-01, 05713-03, 21813-01, 35213-03, 05714-10, 23913-01, 36013-01, 06313-05, 24113-04, 36113-02, 06314-04, 24213-04, and 36413-05. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with mechanical impact.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Tenex Health Inc
- Distributor
- Worldwide Distribution - USA (Nationwide) and Internationally to Panama and Singapore.
- Lot
- Lot Numbers: 00614-03, 06513-02, 24913-01, 00713-06, 06514-02, 25313-01, 00814-02, 06914-01, 25513-02, 01114-02, 07213-04, 25613-01, 01514-03, 07713-18, 26113-06, 01714-04, 08113-04, 26313-04, 01813-02, 08413-13, 26713-04, 02314-08, 09313-05, 26813-01, 02413-04, 10213-07, 27013-10, 02714-02, 10613-01, 27613-02, 02914-01, 11213-07, 28013-05, 03014-02, 11613-06, 28313-02, 03113-02, 12113-02, 29513-03, 03114-02, 12713-03, 29713-01, 03414-03, 13413-01, 30913-07, 03514-05, 14413-05, 31113-08, 03714-03, 15013-04, 31513-02, 03814-06, 15413-02, 31713-05, 04114-02, 15413-03, 32213-01, 04213-01, 16513-04, 32413-01, 04214-05, 17013-02, 32613-01, 04414-03, 17013-09, 32913-07, 04514-02, 17613-08, 33613-09, 04814-07, 17913-16, 33813-05, 05014-02, 18413-05, 34013-03, 05113-01, 18413-06, 34313-01, 05113-02, 18613-03, 34413-04, 05114-09, 19213-03, 34513-04, 05514-01, 19713-04, 34713-08, 05614-04, 20713-06, 35113-01, 05713-03, 21813-01, 35213-03, 05714-10, 23913-01, 36013-01, 06313-05, 24113-04, 36113-02, 06314-04, 24213-04, and 36413-05.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANellcor SpO2 Upgrade kit, consists of a Nellcor SpO2 circuit board, an OEM part manufactured by a third partyMay 14, 2014
- FDAADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIAMay 14, 2014
- FDAPanda® Wall-Mount Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who areMay 14, 2014
- FDABausch & Lomb, Biotrue, Multi-purpose contact lens solution, 10 fl.oz., Sterile.May 14, 2014
- FDAPanda® Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who areMay 14, 2014
- FDANonabsorbable Surgical Suture, U.S.P., Sterile - Do Not Resterilize, Deknatel. Indicated for use in softMay 14, 2014
Questions about this recall
Was TX1 Tissue Removal System. TX1 Procedure Pack P/N 554-1003-001 (outer box), TX1 MicroTip P/N 554-1002-001 (inner tray). For use with the TX System Console as an ultrasonic surgical aspirator of soft tissue. recalled?
Yes. FDA published recall Z-1552-2014 on May 14, 2014. Reported hazard: The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with mechanical impact.
What should I do if I have this product?
Voluntary: Firm initiated · Letter