eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood pressure). For medical use by professional medical staff.
Published May 4, 2016 · Recall initiated: March 10, 2016
Summary
FDA records medical device Class II recall Z-1557-2016, published May 4, 2016 and initiated March 10, 2016. Recalling firm: Philips Visicu; Status: Terminated; Reported quantity: 9; Lot: 453564506091 eCareCoordinator 1.3. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later. An error message is generated and forwarded to operations. Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement. If the pa
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Visicu
- Distributor
- US Distribution -- to the stated of FL, MA, MI, MS, PA, and KS.
- Lot
- 453564506091 eCareCoordinator 1.3
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAev3, PowerCross" 0.018" OTW PTA Dilatation Catheter, model: AB18W030200150, Sterile EO, Rx Only.May 4, 2016
- FDA3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.May 4, 2016
- FDANewport Creamery Butter Crunch Ice Cream, 3Gal Product Code: 35108May 4, 2016
- FDAMagellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen MedicalMay 4, 2016
- FDAEmack & Bolios Space Cake Ice Cream, 3gal Product Code: 36273May 4, 2016
- FDABD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determiningMay 4, 2016
Questions about this recall
Was eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood pressure). For medical use by professional medical staff. recalled?
Yes. FDA published recall Z-1557-2016 on May 4, 2016. Reported hazard: The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later. An error message is generated and forwarded to operations. Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement. If the pa
What should I do if I have this product?
Voluntary: Firm initiated · Letter