FDAFDA · Food and drugs·Recall number Z-1557-2016

eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood pressure). For medical use by professional medical staff.

Published May 4, 2016 · Recall initiated: March 10, 2016

Summary

FDA records medical device Class II recall Z-1557-2016, published May 4, 2016 and initiated March 10, 2016. Recalling firm: Philips Visicu; Status: Terminated; Reported quantity: 9; Lot: 453564506091 eCareCoordinator 1.3. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later. An error message is generated and forwarded to operations. Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement. If the pa

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Visicu
Distributor
US Distribution -- to the stated of FL, MA, MI, MS, PA, and KS.
Lot
453564506091 eCareCoordinator 1.3
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood pressure). For medical use by professional medical staff. recalled?

Yes. FDA published recall Z-1557-2016 on May 4, 2016. Reported hazard: The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later. An error message is generated and forwarded to operations. Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement. If the pa

What should I do if I have this product?

Voluntary: Firm initiated · Letter