BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.
Published May 4, 2016 · Recall initiated: March 31, 2016
Summary
FDA records medical device Class II recall Z-1570-2016, published May 4, 2016 and initiated March 31, 2016. Recalling firm: Becton, Dickinson and Company, BD Biosciences; Status: Terminated; Reported quantity: 1743; 467 US and 1276 International; Lot: Model: 340334; Manufacaturing lot: 534964. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to patient safety or impact to data if this product is used.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton, Dickinson and Company, BD Biosciences
- Distributor
- Worldwide Distribution - US including MD, CA GA. WA. PA. CO. TN. FL. NC. DE. VA, CT, LA. NC, NY, OH, MN, NH and Internationally to China, Japan, and India.
- Lot
- Model: 340334; Manufacaturing lot: 534964
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Multitest 6-Color TBNK CE-IVD, REF: 644611; BD Multitest 6-Color TBNK Kit with BD Trucount Tubes CE-IVD,July 24, 2024
- FDACD11b APC: ASR, REF: 340936, and CE, REF: 333143January 17, 2024
- FDABD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BDNovember 9, 2022
- FDABD Anti-Kappa PE (Cat. No. 346587) is an analyte specific reagent for the Anti-Kappa antibody, whichJune 8, 2022
- FDAMagellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen MedicalMay 4, 2016
- FDAPanorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store,May 4, 2016
Questions about this recall
Was BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader. recalled?
Yes. FDA published recall Z-1570-2016 on May 4, 2016. Reported hazard: One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to patient safety or impact to data if this product is used.
What should I do if I have this product?
Voluntary: Firm initiated · Letter