ABACUS TPN Calculation Software, Product Codes: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition), 8300-0156 ABACUS 3.0 CE, 8300-0157 ABACUS 3.0 SE, 8300-0158 ABACUS 3.0 ME, 8300-0166 ABACUS 3.1 DE, 8300-0167 ABACUS 3.1 CE, 8300-0168 ABACUS 3.1 SE, 8300-0169 ABACUS 3.1 ME. ABACUS is a Windows - based order entry TPN software application for comprehensive calculations and label printing. The primary use of Abacus is for Total Parenteral Nutrition (TPN) order calculations.
Published May 21, 2014 · Recall initiated: April 2, 2014
Summary
FDA records medical device Class I recall Z-1566-2014, published May 21, 2014 and initiated April 2, 2014. Recalling firm: Baxter Corporation Englewood; Status: Terminated; Reported quantity: 2034; Lot: Product Code 8300-0045, 8300-0046, 8300-0047, 8300-0156, 8300-0157, 8300-0158, 8300-0166, 8300-0167, 8300-0168, 8300-0169. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creation of TPN orders. 2. ABACUS v3.1 may automatically add additional sterile water to a formula equal to the volume of a premix resulting in an over dilution. 3. All software versions of ABACUS software display the calcium phosphate curve points for Premasol incorrectly. 4. All software v
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Corporation Englewood
- Distributor
- Worldwide Distribution - USA nationwide, Canada, Colombia, Dominican Republic, Panama, Austria, Bosnia Herzegovina, Bulgaria, Bahrain, Estonia, France, Germany, Georgia, Hungary, Ireland, Israel, Jordan, Kuwait, Latvia, Morocco, Oman, Qatar, Saudi Arabia, Slovenia, Slovakia, South Africa, Switzerland, United Arab Emirates, United Kingdom, Hong Kong, and Taiwan.
- Lot
- Product Code 8300-0045, 8300-0046, 8300-0047, 8300-0156, 8300-0157, 8300-0158, 8300-0166, 8300-0167, 8300-0168, 8300-0169
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA0.2 Micron Filter, 50 mm Product Usage: The 0.2 Micron Filter is a bacteria and particulate filter for aqueousOctober 12, 2016
- FDABaxter, RAPIDFILL Syringe Strip, 10 ml. For use with RAPIDFILL Automated Syringe Filler (ASF)February 10, 2016
- FDASelf-Righting Luer Lock Tip Cap, Green and Self-Righting Luer Slip Tip Cap, Yellow, Baxter HealthcareDecember 31, 2014
- FDAAbacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.July 2, 2014
- FDAPurity Organic - Organic Mango Mango Tommy Atkins CAT 1, Size 07, 08, 10 and 12. Product of Mexico,May 21, 2014
- FDAAlaris Pump model 8100 with software version 9.1.18 and software upgrade kits with "Delay Until" Option andMay 21, 2014
Questions about this recall
Was ABACUS TPN Calculation Software, Product Codes: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition), 8300-0156 ABACUS 3.0 CE, 8300-0157 ABACUS 3.0 SE, 8300-0158 ABACUS 3.0 ME, 8300-0166 ABACUS 3.1 DE, 8300-0167 ABACUS 3.1 CE, 8300-0168 ABACUS 3.1 SE, 8300-0169 ABACUS 3.1 ME. ABACUS is a Windows - based order entry TPN software application for comprehensive calculations and label printing. The primary use of Abacus is for Total Parenteral Nutrition (TPN) order calculations. recalled?
Yes. FDA published recall Z-1566-2014 on May 21, 2014. Reported hazard: 1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creation of TPN orders. 2. ABACUS v3.1 may automatically add additional sterile water to a formula equal to the volume of a premix resulting in an over dilution. 3. All software versions of ABACUS software display the calcium phosphate curve points for Premasol incorrectly. 4. All software v
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail