FDAFDA · Food and drugs·Recall number Z-1847-2014
Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.
Published July 2, 2014 · Recall initiated: May 28, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may display 1,000 times less protein than is actually present in the solution.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Baxter Corporation Englewood
- Distributor
- US Nationwide Distribution and the state of VA.
- Lot
- Product Code: 8300-0167, 8300-0168, 8300-0169
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.