FDAFDA · Food and drugs·Recall number Z-1566-2019

Maquet Getinge-BEQ-TOP 5909 1/4 PACK W QPID Material: 701064992

Published June 5, 2019 · Recall initiated: April 18, 2019

Summary

FDA records medical device Class II recall Z-1566-2019, published June 5, 2019 and initiated April 18, 2019. Recalling firm: Maquet Cardiovascular, LLC; Status: Terminated; Reported quantity: N/A; Lot: Batch Numbers: 3000043679 3000047252 3000050833 3000053134 3000054956 3000056048 3000060481 3000063710 3000067185 3000071400 3000073429 3000075387 3000075899. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Maquet Cardiovascular, LLC
Distributor
Nationwide
Lot
Batch Numbers: 3000043679 3000047252 3000050833 3000053134 3000054956 3000056048 3000060481 3000063710 3000067185 3000071400 3000073429 3000075387 3000075899
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Maquet Getinge-BEQ-TOP 5909 1/4 PACK W QPID Material: 701064992 recalled?

Yes. FDA published recall Z-1566-2019 on June 5, 2019. Reported hazard: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

What should I do if I have this product?

Voluntary: Firm initiated · Letter