Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.

Published October 1, 2025 · Recall initiated: August 6, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
Maquet Cardiovascular, LLC
Distributor
US Nationwide distribution in the states of FL, IL, MD & TX.
Lot
Model No. C-VH-6000; UDI: 00607567701274; Lot/Serial No. 3000486429, 3000487231, 3000487726, 3000488639.
Risk classification
Class II