Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.
Published October 1, 2025 · Recall initiated: August 6, 2025
Summary
FDA records medical device Class II recall Z-2636-2025, published October 1, 2025 and initiated August 6, 2025. Recalling firm: Maquet Cardiovascular, LLC; Status: Ongoing; Reported quantity: 98 units; Lot: Model No. C-VH-6000; UDI: 00607567701274; Lot/Serial No. 3000486429, 3000487231, 3000487726, 3000488639. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Cardiovascular, LLC
- Distributor
- US Nationwide distribution in the states of FL, IL, MD & TX.
- Lot
- Model No. C-VH-6000; UDI: 00607567701274; Lot/Serial No. 3000486429, 3000487231, 3000487726, 3000488639.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile,July 22, 2026
- FDAEntecavir Tablets, USP, 0.5 mg, 30-count bottle, Rx Only, Manufactured by: Zydus Lifesciences Ltd., Ahmedabad,October 8, 2025
- FDAAllura Xper FD10 OR Table Product Codes: (1)722022, (2) 722033; Associated 510(k) numbers: K102005, K130842,October 1, 2025
- FDAAcrobat-i Positioner. Model Number C-XP-5000ZNovember 27, 2024
- FDAAcrobat SUV Vacuum Stabilizer System, ST. Model Number C-OM-9000ZNovember 27, 2024
- FDAAcrobat-i Vacuum Stabilizer System. Model Number C-OM-10000ZNovember 27, 2024
Questions about this recall
Was Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy. recalled?
Yes. FDA published recall Z-2636-2025 on October 1, 2025. Reported hazard: Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone