MedStream Programmable Infusion Pump, 20 mL, Product Code: 91-4200 The MedStream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intrathecal delivery of Morphine or Baclofen in the EMEA market.

Published May 21, 2014 · Recall initiated: November 10, 2015

Summary

FDA records medical device Class II recall Z-1571-2014, published May 21, 2014 and initiated November 10, 2015. Recalling firm: Codman & Shurtleff, Inc.; Status: Terminated; Reported quantity: 90; Lot: Serial Numbers: NNBLTL, NNBLT5, NNBLR9, NNBLT2, NNBLJP, NNBLL2, NNBLJR, NNBLLH, NNBLLJ, NNBLLG, NNBLT0, NNBLT1, NNBLL7, NNBLTC, NNBLTN, NNBLTK, NPBLZ7, NPBLZH, NPBLZP, NPBLZN, NPBLZM, NPBLZK, NPBLZ4, NPBLZJ, NPBLY3, NPBLY5, NPBLY4, NPBLY7, NPBLZT, NPBLY9, NPBLZR, NPBL7M, NPBL74, NPBMFN, NPBMF0, NPBMHJ, NPBMHK, NPBL71 NPBL9H, NNBLLP, NNBLLN, NNBLLC, NNBLLD, NNBLL3, NPBL0F, NNBLTH, NPBL6N, NPBL6M, NPBL6L, NPBL1T, NPBL17, NPBL1Y, NPBL1W, NPBL75, NPBL70, NPBL76, NNBLG2, NPBMFG, NPBMF4, NPBMF2, NPBMF1, NPBMF5, NPBMH2, NPBMH0, NNBLJM, NNBLJK, NNBLJN, NNBLHZ, NNBLHW, NNBLH2, NNBLH1, NNBLH0, NNBLTB, NNBLL4, NPBL1R, NPBL1P, NPBL1N, NPBL1L, NPBL1J, NPBL9N, NPBL9K, NPBL9L, NNBLHN, NNBLGV, NNBLHT, NNBLHM, NNBLG0, NNBLLK, NMBKBW, NMBJ5G. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A small number of specific pumps may trigger a low battery alarm at the beginning of the implanted life.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Codman & Shurtleff, Inc.
Distributor
Worldwide Distribution including the countries of Belgium Germany, Spain, United Kingdom, Ireland, Italy, Russia and Turkey.
Lot
Serial Numbers: NNBLTL, NNBLT5, NNBLR9, NNBLT2, NNBLJP, NNBLL2, NNBLJR, NNBLLH, NNBLLJ, NNBLLG, NNBLT0, NNBLT1, NNBLL7, NNBLTC, NNBLTN, NNBLTK, NPBLZ7, NPBLZH, NPBLZP, NPBLZN, NPBLZM, NPBLZK, NPBLZ4, NPBLZJ, NPBLY3, NPBLY5, NPBLY4, NPBLY7, NPBLZT, NPBLY9, NPBLZR, NPBL7M, NPBL74, NPBMFN, NPBMF0, NPBMHJ, NPBMHK, NPBL71 NPBL9H, NNBLLP, NNBLLN, NNBLLC, NNBLLD, NNBLL3, NPBL0F, NNBLTH, NPBL6N, NPBL6M, NPBL6L, NPBL1T, NPBL17, NPBL1Y, NPBL1W, NPBL75, NPBL70, NPBL76, NNBLG2, NPBMFG, NPBMF4, NPBMF2, NPBMF1, NPBMF5, NPBMH2, NPBMH0, NNBLJM, NNBLJK, NNBLJN, NNBLHZ, NNBLHW, NNBLH2, NNBLH1, NNBLH0, NNBLTB, NNBLL4, NPBL1R, NPBL1P, NPBL1N, NPBL1L, NPBL1J, NPBL9N, NPBL9K, NPBL9L, NNBLHN, NNBLGV, NNBLHT, NNBLHM, NNBLG0, NNBLLK, NMBKBW, NMBJ5G
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MedStream Programmable Infusion Pump, 20 mL, Product Code: 91-4200 The MedStream programmable pump is intended for the intrathecal delivery of Baclofen in the US market and for the intrathecal delivery of Morphine or Baclofen in the EMEA market. recalled?

Yes. FDA published recall Z-1571-2014 on May 21, 2014. Reported hazard: A small number of specific pumps may trigger a low battery alarm at the beginning of the implanted life.

What should I do if I have this product?

Voluntary: Firm initiated · Letter