FDAFDA · Food and drugs·Recall number Z-1588-2013
Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems. Produce cross-sectional images of the body by computer reconstruction of x-ray transmission.
Published July 3, 2013 · Recall initiated: March 21, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information page is the patient name selected with the mouse. Also, it is possible to begin a normal clinical scan while the anti-virus scan is running. However, if the user is running a bol
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Nationwide Distribution-including the states of AR, AZ, DE, IN, KS, LA, MA, MN, MS, MT, NJ, NY, NY, OH, OR, PA, TN, TX, VT, WV, and WY.
- Lot
- The following codes are affected by this recall:System Code #728231, Serial #: 9654, 91003, 95130, 95157, 95504 System Code #728326, Serial #: 32000,300003,300004,300005,300010,300011,300012,300013, 300014,300015,300018, 300027,300041,300053,300054, 300059, 300066, 300070,300071, 300073, 300074, 30077, 300078, 300079, 300080,300082,300084,300086,300091,300092, 300095, 300101, 300102,300112, 300119, 300127, 300128, 300129, 300131, 300132, 310064, 320030 System Code #728321, Serial #: 310096, 310070 System Code #728323, Serial #: 320033
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States