FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 37303-20, 37303-22, 37303-24, 37303-26, 37303-28, 37303-30. 2. FLS - Foot Locking Plates System, Cortical Screw; Article Numbers: 32271-10, 32271-12, 32271-14, 32271-16, 32271-18, 32271-20, 32271-22, 32271-24, 32271-26, 32271-28, 32271-32, 32271-34. 3. FLS - Foot Locking Plates System, Cortical Stabilization Screw; Article Numbers: 37304-10, 37304-12, 37304-14, 37304-16, 37304-18, 37304-20, 37304-22, 37304-24. 4. FLS - Foot Locking Plates System, FLS Plate, Curved; Article Numbers: 21413-6. 5. FLS - Foot Locking Plates System, FLS Plate, H-Shape; Article Numbers: 21431-ME. 6. FLS - Foot Locking Plates System, FLS Plate, L-Shape; Article Numbers: 21482, 21483. 7. FLS - Foot Locking Plates System, FLS Plate, Straight; Article Numbers: 21421-5. 8. FLS - Foot Locking Plates System, Stabilization Screw; Article Numbers: 37241-10, 37241-12, 37241-14, 37241-16, 37241-18, 37241-20, 37241-22, 37241-24, 37241-26, 37241-28, 37241-30. 9. HOL - Hallux Osteotomy Locking Plate, Hallux Osteotomy Plate; Article Numbers: 21015-08, 21015-09.
Published April 1, 2026 · Recall initiated: January 23, 2026
Summary
FDA records medical device Class II recall Z-1605-2026, published April 1, 2026 and initiated January 23, 2026. Recalling firm: I.T.S. GmbH; Status: Ongoing; Reported quantity: 15,517 units; Lot: 1. Article Number (Lot Numbers): 37303-10 (18/151720, 18/268729, 679/032432, 679/072268, 679/102237, 679/102307), 37303-12 (18/570637, 679/032434, 679/072269, 679/102346, 679/112227), 37303-14 (18/751091, 26/F66913, 679/012510, 679/062288, 679/102359, 679/112410), 37303-16 (679/072263, 679/102245, 18/151841, 26/F66914, 18/331193, 18/543625, 679/012511, 679/042403, 679/102360), 37303-18 (18/151942, 18/210467, 18/231415, 18/341194, 18/572633, 18/771093, 679/012512, 679/022507, 679/042404, 679/102248, 679/112319), 37303-20 (18/152043, 18/211468, 18/351195, 18/573634, 18/701585, 18/781094, 679/022508, 679/042420, 679/112216, 679/112328), 37303-22 (18/171705, 18/212469, 18/361196, 18/574635, 18/791095, 679/022509, 679/042417, 679/112224, 679/112340), 37303-24 (18/575639, 18/741572, 26/F66915, 679/022510, 679/112226, 679/122313, 679/122333), 37303-26 (18/152346, 18/214471, 26/F66916, 679/012403, 679/112230, 679/122323), 37303-28 (26/F66917, 679/112231, 679/122328), 37303-30 (18/192716, 18/591743, 679/122334). 2. Article Number (Lot Numbers): 32271-10 (18/061949, 18/231167, 679/062351), 32271-12 (18/048927, 26/F66235, 679/042215, 679/062342, 679/062405, 679/062489, 679/072271), 32271-14 (18/049928, 18/231165, 26/F66140, 679/042210, 679/052340, 679/052518, 679/062471, 679/062490, 679/072267), 32271-16 (18/231164, 18/311125, 26/F66141, 679/042216, 679/052519, 679/062315, 679/062483, 679/072272, 679/072401), 32271-18 (18/062950, 26/F65556, 26/F66236, 679/032261, 679/062332, 679/072402), 32271-20 (18/063951, 26/F66237, 679/052326), 32271-22 (18/064952, 26/F65557, 679/032263, 679/062354), 32271-24 (18/065953, 26/F65558, 679/062333), 32271-26 (679/032265), 32271-28 (26/F65559, 679/062357, 679/052520), 32271-32 (26/F65561), 32271-34 (26/F65562). 3. Article Number (Lot Numbers): 37304-10 (18/635486, 26/F24799, 679/012419, 86/011176), 37304-12 (18/061352, 18/521172, 18/5211721, 18/576643, 18/630487, 679/012433, 679/092207, 679/122409), 37304-14 (18/081360, 18/531173, 18/5311731, 18/577644, 18/631488, 679/012434, 679/092208, 679/122410), 37304-16 (18/541174, 18/578645, 18/632489, 679/012444, 679/092201, 679/122411), 37304-18 (18/092020, 18/551175, 18/633490, 679/022403, 679/092202), 37304-20 (18/011320, 18/561176, 18/634491, 679/012287, 679/012420, 679/082217, 86/483331), 37304-22 (18/241273, 679/012288, 679/012421), 37304-24 (18/021321, 18/571177, 679/012289, 679/012445). 4. Article Number (Lot Numbers): 21413-6 (26/F52124). 5. Article Number (Lot Numbers): 21431-ME (679/102045). 6. Article Number (Lot Numbers): 21482 (679/092020), 21483 (26/F55902). 7. Article Number (Lot Numbers): 21421-5 (26/F65377, 679/092039). 8. Article Number (Lot Numbers): 37241-10 (18/012169, 18/610104, 18/701013), 37241-12 (18/231295, 18/767108), 37241-14 (18/511016, 18/537392, 26/F24807, 679/012280), 37241-16 (18/005340, 18/121106, 18/231387, 18/521017, 18/560517), 37241-18 (18/232328, 18/531018, 679/072408, 86/311848), 37241-20 (18/233388, 18/541020, 18/562519, 679/072419, 86/287847), 37241-22 (18/066109, 18/234389, 18/551021, 26/F24594, 26/F24811), 37241-24 (18/561022, 18/564521, 18/608110, 26/F24595), 37241-26 (18/611111, 18/711023, 26/F24596, 26/F24813), 37241-28 (18/612112, 18/721024, 18/811597, 26/F24597, 86/485508), 37241-30 (18/015172, 18/613113, 18/731019, 26/F24598). 9. Article Number (Lot Numbers): 21015-08 (00/0123-12, 00/0321-43, 00/0723-07), 21015-09 (00/0123-13). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- I.T.S. GmbH
- Distributor
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Lot
- 1. Article Number (Lot Numbers): 37303-10 (18/151720, 18/268729, 679/032432, 679/072268, 679/102237, 679/102307), 37303-12 (18/570637, 679/032434, 679/072269, 679/102346, 679/112227), 37303-14 (18/751091, 26/F66913, 679/012510, 679/062288, 679/102359, 679/112410), 37303-16 (679/072263, 679/102245, 18/151841, 26/F66914, 18/331193, 18/543625, 679/012511, 679/042403, 679/102360), 37303-18 (18/151942, 18/210467, 18/231415, 18/341194, 18/572633, 18/771093, 679/012512, 679/022507, 679/042404, 679/102248, 679/112319), 37303-20 (18/152043, 18/211468, 18/351195, 18/573634, 18/701585, 18/781094, 679/022508, 679/042420, 679/112216, 679/112328), 37303-22 (18/171705, 18/212469, 18/361196, 18/574635, 18/791095, 679/022509, 679/042417, 679/112224, 679/112340), 37303-24 (18/575639, 18/741572, 26/F66915, 679/022510, 679/112226, 679/122313, 679/122333), 37303-26 (18/152346, 18/214471, 26/F66916, 679/012403, 679/112230, 679/122323), 37303-28 (26/F66917, 679/112231, 679/122328), 37303-30 (18/192716, 18/591743, 679/122334). 2. Article Number (Lot Numbers): 32271-10 (18/061949, 18/231167, 679/062351), 32271-12 (18/048927, 26/F66235, 679/042215, 679/062342, 679/062405, 679/062489, 679/072271), 32271-14 (18/049928, 18/231165, 26/F66140, 679/042210, 679/052340, 679/052518, 679/062471, 679/062490, 679/072267), 32271-16 (18/231164, 18/311125, 26/F66141, 679/042216, 679/052519, 679/062315, 679/062483, 679/072272, 679/072401), 32271-18 (18/062950, 26/F65556, 26/F66236, 679/032261, 679/062332, 679/072402), 32271-20 (18/063951, 26/F66237, 679/052326), 32271-22 (18/064952, 26/F65557, 679/032263, 679/062354), 32271-24 (18/065953, 26/F65558, 679/062333), 32271-26 (679/032265), 32271-28 (26/F65559, 679/062357, 679/052520), 32271-32 (26/F65561), 32271-34 (26/F65562). 3. Article Number (Lot Numbers): 37304-10 (18/635486, 26/F24799, 679/012419, 86/011176), 37304-12 (18/061352, 18/521172, 18/5211721, 18/576643, 18/630487, 679/012433, 679/092207, 679/122409), 37304-14 (18/081360, 18/531173, 18/5311731, 18/577644, 18/631488, 679/012434, 679/092208, 679/122410), 37304-16 (18/541174, 18/578645, 18/632489, 679/012444, 679/092201, 679/122411), 37304-18 (18/092020, 18/551175, 18/633490, 679/022403, 679/092202), 37304-20 (18/011320, 18/561176, 18/634491, 679/012287, 679/012420, 679/082217, 86/483331), 37304-22 (18/241273, 679/012288, 679/012421), 37304-24 (18/021321, 18/571177, 679/012289, 679/012445). 4. Article Number (Lot Numbers): 21413-6 (26/F52124). 5. Article Number (Lot Numbers): 21431-ME (679/102045). 6. Article Number (Lot Numbers): 21482 (679/092020), 21483 (26/F55902). 7. Article Number (Lot Numbers): 21421-5 (26/F65377, 679/092039). 8. Article Number (Lot Numbers): 37241-10 (18/012169, 18/610104, 18/701013), 37241-12 (18/231295, 18/767108), 37241-14 (18/511016, 18/537392, 26/F24807, 679/012280), 37241-16 (18/005340, 18/121106, 18/231387, 18/521017, 18/560517), 37241-18 (18/232328, 18/531018, 679/072408, 86/311848), 37241-20 (18/233388, 18/541020, 18/562519, 679/072419, 86/287847), 37241-22 (18/066109, 18/234389, 18/551021, 26/F24594, 26/F24811), 37241-24 (18/561022, 18/564521, 18/608110, 26/F24595), 37241-26 (18/611111, 18/711023, 26/F24596, 26/F24813), 37241-28 (18/612112, 18/721024, 18/811597, 26/F24597, 86/485508), 37241-30 (18/015172, 18/613113, 18/731019, 26/F24598). 9. Article Number (Lot Numbers): 21015-08 (00/0123-12, 00/0321-43, 00/0723-07), 21015-09 (00/0123-13).
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Austria
Related recalls
- FDAHCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10,April 1, 2026
- FDAOlecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1.April 1, 2026
- FDAFlexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgicalApril 1, 2026
- FDAPhilips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.April 1, 2026
- FDADoraditas de Azucar UPC 5901234123457April 1, 2026
- FDAPhilips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.April 1, 2026
Questions about this recall
Was FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 37303-20, 37303-22, 37303-24, 37303-26, 37303-28, 37303-30. 2. FLS - Foot Locking Plates System, Cortical Screw; Article Numbers: 32271-10, 32271-12, 32271-14, 32271-16, 32271-18, 32271-20, 32271-22, 32271-24, 32271-26, 32271-28, 32271-32, 32271-34. 3. FLS - Foot Locking Plates System, Cortical Stabilization Screw; Article Numbers: 37304-10, 37304-12, 37304-14, 37304-16, 37304-18, 37304-20, 37304-22, 37304-24. 4. FLS - Foot Locking Plates System, FLS Plate, Curved; Article Numbers: 21413-6. 5. FLS - Foot Locking Plates System, FLS Plate, H-Shape; Article Numbers: 21431-ME. 6. FLS - Foot Locking Plates System, FLS Plate, L-Shape; Article Numbers: 21482, 21483. 7. FLS - Foot Locking Plates System, FLS Plate, Straight; Article Numbers: 21421-5. 8. FLS - Foot Locking Plates System, Stabilization Screw; Article Numbers: 37241-10, 37241-12, 37241-14, 37241-16, 37241-18, 37241-20, 37241-22, 37241-24, 37241-26, 37241-28, 37241-30. 9. HOL - Hallux Osteotomy Locking Plate, Hallux Osteotomy Plate; Article Numbers: 21015-08, 21015-09. recalled?
Yes. FDA published recall Z-1605-2026 on April 1, 2026. Reported hazard: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail