FDAFDA · Food and drugs·Recall number Z-1606-2026

HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90.

Published April 1, 2026 · Recall initiated: January 23, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
I.T.S. GmbH
Distributor
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Lot
Article Number (Lot Numbers): 31207-10 (679/112002), 31207-12 (679/092133, 679/102343, 679/112006), 31207-14 (679/102050), 31207-16 (679/102345, 679/102049), 31207-18 (679/0323110, 679/102355, 679/112001), 31207-20 (679/0323111, 679/102051, 679/102356), 31207-24 (679/102028), 31207-26 (679/112008), 31207-28 (679/112009), 31207-30 (679/0323140, 86/483377), 31257-12 (679/081904), 31257-14 (679/042323, 679/081905), 31257-16 (679/042324), 31257-18 (679/112308, 679/022128), 31257-20 (679/022123, 679/042326, 679/112321), 31257-22 (679/022124, 679/042327, 679/112325), 31257-24 (679/022125, 679/112326), 31257-28 (679/022130, 679/052306), 31257-30 (679/022131), 31307-18 (679/061907), 31307-32 (679/062285), 31307-34 (679/062286), 31307-36 (679/062287), 31307-38 (679/032118), 31358-30 (26/F67790), 31437-30 (679/071916), 31437-38 (679/071918), 31437-40 (679/032406, 679/071903), 31437-50 (679/071920), 31757-100 (35/51892), 31757-120 (86/320792), 31757-45 (86/320793), 31757-80 (26/F55872), 31757-85 (26/F55873), 31757-90 (26/F55874).
Risk classification
Class II
Area (FDA)
device
Country
Austria