Medtronic REF: MMT-7020C Guardian Sensor (3) kit for the following packaging: GS3 SEMI 5PK OUS/ GTIN: 20763000192086, 5PK SF NEW GS3 OUS/GTIN: 20763000358451 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.
Published May 26, 2021 · Recall initiated: February 26, 2021
Summary
FDA records medical device Class II recall Z-1606-2021, published May 26, 2021 and initiated February 26, 2021. Recalling firm: Medtronic Inc.; Status: Ongoing; Reported quantity: 497,747 units; Lot: All Lot numbers within active shelf life. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Inc.
- Distributor
- Worldwide distribution.
- UPC
- 20763000192086
- Lot
- All Lot numbers within active shelf life
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090September 25, 2024
- FDACOBALT XT VR MRI SureScan, Model Number DVPA2D1, Implantable Cardioverter DefibrillatorJuly 3, 2024
- FDACROME VR MRI SureScan, Model Number DVPC3D1, Implantable Cardioverter DefibrillatorJuly 3, 2024
- FDACobalt XT DR MRI SureScan, Model Number DDPA2D1, Implantable Cardioverter DefibrillatorJuly 3, 2024
- FDAUdi Gluten Free Double Chocolate Muffins, 10oz (284g) retail container of 4 muffins, 6 retail units perMay 26, 2021
- FDABioseal, Central Line Dressing Kit, containing ChloraPREP Applicator 3 mL, REF: CVC009/20, Sterile EO, SingleMay 26, 2021
Questions about this recall
Was Medtronic REF: MMT-7020C Guardian Sensor (3) kit for the following packaging: GS3 SEMI 5PK OUS/ GTIN: 20763000192086, 5PK SF NEW GS3 OUS/GTIN: 20763000358451 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US. recalled?
Yes. FDA published recall Z-1606-2021 on May 26, 2021. Reported hazard: Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
How do I know if my unit is part of the recall?
Compare your product details with the notice. UPC: 20763000192086.