ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.

Published May 21, 2014 · Recall initiated: April 22, 2014

Summary

FDA records medical device Class II recall Z-1615-2014, published May 21, 2014 and initiated April 22, 2014. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 8 devices; Lot: Catalog Number: 11-210062 Lot Number Identification: 972230. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens, a revision surgery may be necessary.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biomet, Inc.
Distributor
US Nationwide Distribution in the states of NY, MD, NC, SC, MI, OR, and NV.
Lot
Catalog Number: 11-210062 Lot Number Identification: 972230
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant. recalled?

Yes. FDA published recall Z-1615-2014 on May 21, 2014. Reported hazard: This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens, a revision surgery may be necessary.

What should I do if I have this product?

Voluntary: Firm initiated · Letter