ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip; 51 count MedBen strip 51 count Mail Order strip; Sample Kit 16 per case. The ACCU-CHEK Compact Plus Test Strips are for use with the ACCU-CHEK Compact Plus Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm.
Published May 21, 2014 · Recall initiated: April 14, 2014
Summary
FDA records medical device Class II recall Z-1616-2014, published May 21, 2014 and initiated April 14, 2014. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 590,136 boxes; Lot: 5599423160, 5599415160, 5919967160, 5919959160, 5919983160, 5907675001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Distributor
- Nationwide Distribution.
- Lot
- 5599423160, 5599415160, 5919967160, 5919959160, 5919983160, 5907675001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAcobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, inNovember 27, 2024
- FDACreatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitativeOctober 2, 2024
- FDAAlbumin Gen.2, Catalog 05166861190October 21, 2020
- FDAElecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virusDecember 18, 2019
- FDAExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.May 21, 2014
- FDAMcKesson Radiology-PACS. McKesson Radiology is a medical image and information application that is intended toMay 21, 2014
Questions about this recall
Was ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip; 51 count MedBen strip 51 count Mail Order strip; Sample Kit 16 per case. The ACCU-CHEK Compact Plus Test Strips are for use with the ACCU-CHEK Compact Plus Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm. recalled?
Yes. FDA published recall Z-1616-2014 on May 21, 2014. Reported hazard: Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit