48" Lead Wires, Replacement Part Number: 1067724-4
Published May 1, 2024 · Recall initiated: March 22, 2024
Summary
FDA records medical device Class II recall Z-1619-2024, published May 1, 2024 and initiated March 22, 2024. Recalling firm: EBI, LLC; Status: Ongoing; Reported quantity: 594 units (2 lead wires assembly); Lot: UDI-DI: 00812301020324 All devices distributed since May 1, 2023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- EBI, LLC
- Distributor
- US Nationwide distribution.
- Lot
- UDI-DI: 00812301020324 All devices distributed since May 1, 2023
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABiomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an establishedMay 1, 2024
- FDA20" Lead Wires, Replacement Part Number: 1067724-2May 1, 2024
- FDABiomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an establishedMay 1, 2024
- FDAMariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF:May 1, 2024
- FDATRIDENTII HEMI CLUSTER44B- Indicated for use in primary and revision patients at high risk of hip dislocationMay 1, 2024
- FDATRIDENTII HEMI CLUSTER48D-indicated for use in primary and revision patients at high risk of hip dislocationMay 1, 2024
Questions about this recall
Was 48" Lead Wires, Replacement Part Number: 1067724-4 recalled?
Yes. FDA published recall Z-1619-2024 on May 1, 2024. Reported hazard: Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
What should I do if I have this product?
Voluntary: Firm initiated · Letter