Back to search
FDAFDA · Food and drugs·Recall number Z-1620-2024

Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

Published May 1, 2024 · Recall initiated: February 28, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
SEASPINE ORTHOPEDICS CORPORATION
Distributor
US: CA, FL, CO, TX
Lot
REF/UDI-DI/Lots: MX1-001010/10889981298647/ AW161319D, AW164195D, AW166994D, AW167309D, AW167422D, AW169350F, AW169351F. AMRDXMIS/10889981303297/ AMRDXMIS-001, AMRDXMIS-002, AMRDXMIS-003, AMRDXMIS-004, AMRDXMIS-005
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.