Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.
Published May 1, 2024 · Recall initiated: February 28, 2024
Summary
FDA records medical device Class II recall Z-1620-2024, published May 1, 2024 and initiated February 28, 2024. Recalling firm: SEASPINE ORTHOPEDICS CORPORATION; Status: Ongoing; Reported quantity: 444 Screw Heads; Lot: REF/UDI-DI/Lots: MX1-001010/10889981298647/ AW161319D, AW164195D, AW166994D, AW167309D, AW167422D, AW169350F, AW169351F. AMRDXMIS/10889981303297/ AMRDXMIS-001, AMRDXMIS-002, AMRDXMIS-003, AMRDXMIS-004, AMRDXMIS-005. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- SEASPINE ORTHOPEDICS CORPORATION
- Distributor
- US: CA, FL, CO, TX
- Lot
- REF/UDI-DI/Lots: MX1-001010/10889981298647/ AW161319D, AW164195D, AW166994D, AW167309D, AW167422D, AW169350F, AW169351F. AMRDXMIS/10889981303297/ AMRDXMIS-001, AMRDXMIS-002, AMRDXMIS-003, AMRDXMIS-004, AMRDXMIS-005
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12September 30, 2026
- FDASeaSpine Polyaxial Head REF Number: 41-3010July 29, 2026
- FDABrand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/CatalogJanuary 1, 2025
- FDABrand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 ProductJuly 24, 2024
- FDA48" Lead Wires, Replacement Part Number: 1067724-4May 1, 2024
- FDATRIDENTII HEMI CLUSTER44B- Indicated for use in primary and revision patients at high risk of hip dislocationMay 1, 2024
Questions about this recall
Was Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems. recalled?
Yes. FDA published recall Z-1620-2024 on May 1, 2024. Reported hazard: Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail