CDI Blood Parameter Monitoring System; Catalog Numbers or Part Numbers - 145891, 500AVHCT; Serial Numbers - BPM SN B015552 (installed into CDI 500 Monitor SN 3535)

Published May 28, 2014 · Recall initiated: May 9, 2014

Summary

FDA records medical device Class II recall Z-1634-2014, published May 28, 2014 and initiated May 9, 2014. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 1; Lot: Catalog Number or Part Number 145891 500AVHCT Serial Number: BPM SN 8015552 (installed into CDI 500 Monitor SN 3535). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate, there is a potential risk of unnecesary blood transfusion with resulting risk of infection or inflammatory reaction. BPM shipped without meeting manufacturing accep

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Terumo Cardiovascular Systems Corporation
Distributor
one customer in IA
Lot
Catalog Number or Part Number 145891 500AVHCT Serial Number: BPM SN 8015552 (installed into CDI 500 Monitor SN 3535)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was CDI Blood Parameter Monitoring System; Catalog Numbers or Part Numbers - 145891, 500AVHCT; Serial Numbers - BPM SN B015552 (installed into CDI 500 Monitor SN 3535) recalled?

Yes. FDA published recall Z-1634-2014 on May 28, 2014. Reported hazard: There is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate, there is a potential risk of unnecesary blood transfusion with resulting risk of infection or inflammatory reaction. BPM shipped without meeting manufacturing accep