FDAFDA · Food and drugs·Recall number Z-1636-2016
ABG-HM-1 Hummi Micro Draw Blood Transfer Device Product Usage: It is used for blood transfer and collection from Peripheral Arterial Line Catheters.
Published June 1, 2016 · Recall initiated: October 1, 2015
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Hummingbird Med Devices, Inc. is recalling ABG-HM-1 Hummi Micro Draw Blood Transfer because the connection between the "Hummi" Y connector and the yellow cannula hub may not be sufficiently secure enough and could separate while in use.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Hummingbird Med
- Distributor
- Distributed in the states of CA, KY, MD, GA, and IL.
- Lot
- Lot 15180, 15286, 15287, 15300, 15305
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.