ADVIA Centaur Systems BRAHMS PCT 100 test Sold only OUS; Catalog Number: 10378883.
Published June 4, 2014 · Recall initiated: April 10, 2014
Summary
FDA records medical device Class II recall Z-1659-2014, published June 4, 2014 and initiated April 10, 2014. Recalling firm: Siemens Healthcare Diagnostics, Inc; Status: Terminated; Reported quantity: 16,302 kits; Lot: Kit lots ending in/Exp Date: 020 11 July 2014 021 15 December 2014 022 24 February 2015 023 25 February 2015 024 03 July 2015. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Samples containing fluorescein may show interference with the ADVIA Centaur® Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Afghanistan Algeria Armenia Austria Bahrain Belgium Bulgaria Burkina Faso Canary Islands Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iceland Iran Iraq Ireland Israel Italy Jordan Kazakhstan Latvia Lebanon Lithuania Luxembourg Macedonia Malta Netherlands Norway Oman Pakistan Poland Portugal Qatar Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia Spain Sweden Switzerland Turkey Unit.Arab Emir. Vatican Yemen.
- Lot
- Kit lots ending in/Exp Date: 020 11 July 2014 021 15 December 2014 022 24 February 2015 023 25 February 2015 024 03 July 2015
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASiemens ADVIA Centaur Folate 100 test kit-for IVD of folate in serum or red blood cells SMN:10310308January 29, 2020
- FDASiemensADVIA Centaur aTG (500 test kit) Siemens Material Number (SMN): 10492399May 8, 2019
- FDAAtellica IM aTG (500 test kit) Siemens Material Number (SMN): 10995462May 8, 2019
- FDASiemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461May 8, 2019
- FDAADVIA Centaur Systems VitD Total 100 test, Ready Pack, For OUS; SMN 10699201. Intended for the invitroJune 4, 2014
- FDALIFEPAK 1000 defibrillator. Affected Part Numbers: 320371500XXX The LP1000 (in AED mode) is indicated for useJune 4, 2014
Questions about this recall
Was ADVIA Centaur Systems BRAHMS PCT 100 test Sold only OUS; Catalog Number: 10378883. recalled?
Yes. FDA published recall Z-1659-2014 on June 4, 2014. Reported hazard: Samples containing fluorescein may show interference with the ADVIA Centaur® Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
What should I do if I have this product?
Voluntary: Firm initiated · Letter