What's the hazard
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- IHB OPERATIONS B.V.
- Distributor
- US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
- Lot
- UDI/DI 10841736116071, All Serial Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands