Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Published April 30, 2025 · Recall initiated: March 27, 2025
Summary
FDA records medical device Class II recall Z-1663-2025, published April 30, 2025 and initiated March 27, 2025. Recalling firm: IHB OPERATIONS B.V.; Status: Ongoing; Reported quantity: 112 units; Lot: UDI/DI 10841736116071, All Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- IHB OPERATIONS B.V.
- Distributor
- US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
- Lot
- UDI/DI 10841736116071, All Serial Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDATirofiban Hydrochloride Injection 5 mg/100 mL (50 mcg/mL), 100 mL single-dose container (bag), Rx only,May 7, 2025
- FDATirofiban Hydrochloride Injection 12.5 mg/250 mL (50 mcg/mL), 250 mL single-dose container (bag), Rx only,May 7, 2025
- FDABrand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: AllApril 30, 2025
- FDAThe Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included inApril 30, 2025
- FDAInPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA,April 30, 2025
- FDAZimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +0 Neck, Model Number 802802202; Prosthesis, Hip,April 30, 2025
Questions about this recall
Was Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount recalled?
Yes. FDA published recall Z-1663-2025 on April 30, 2025. Reported hazard: The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail