Sovereign® Compact Phacoemulsification System, v 5.3, Part Number: CMP680300. The system is used for cataract surgery.
Published July 17, 2013 · Recall initiated: May 23, 2013
Summary
FDA records medical device Class II recall Z-1670-2013, published July 17, 2013 and initiated May 23, 2013. Recalling firm: Abbott Medical Optics, Inc.; Status: Terminated; Reported quantity: 534 systems; Lot: Product Name: Sovereign® Compact Console Universal, Part Number: CMP680300. The Sovereign® Compact console is identified with a unique, sequentially assigned 3 digit number identified below as XXX. For Sovereign® Compact systems manufactured at the Third Party Manufacturer, Sanmina-SCI Corporation, the serial number format is YYYY35XXX where Y is for the year manufactured and "35" designates manufacturing at Sanmina-SCI Corporation. The Sanmina-SCI Corporation started manufacturing the Sovereign® Compact v 5.3 systems in February 2012. For Sovereign® Compact systems manufactured at the AMO Manufacturing USA, LLC facility in Milpitas, CA the serial number format is YYYY80XXX where Y is for the year manufactured and "80" designates manufacturing at Milpitas. The Milpitas facility started manufacturing the v 5.3 systems in February 2013. The serial number is contained in the Product Identification Label (part number Z352548) that is attached to the rear of the console for systems manufactured. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Abbott Medical Optics Inc. (AMO) initiated this Urgent Notice of Medical Device Correction and Advisory Notice to all customers who use the AMO Sovereign® Compact Phacoemulsification System, Version 5.3, because it has the potential to lose the vacuum tone mode during surgery.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Medical Optics, Inc.
- Distributor
- Worldwide Distribution, including Nationwide (US) and countries in Latin America, EMEA, and APAC.
- Lot
- Product Name: Sovereign® Compact Console Universal, Part Number: CMP680300. The Sovereign® Compact console is identified with a unique, sequentially assigned 3 digit number identified below as XXX. For Sovereign® Compact systems manufactured at the Third Party Manufacturer, Sanmina-SCI Corporation, the serial number format is YYYY35XXX where Y is for the year manufactured and "35" designates manufacturing at Sanmina-SCI Corporation. The Sanmina-SCI Corporation started manufacturing the Sovereign® Compact v 5.3 systems in February 2012. For Sovereign® Compact systems manufactured at the AMO Manufacturing USA, LLC facility in Milpitas, CA the serial number format is YYYY80XXX where Y is for the year manufactured and "80" designates manufacturing at Milpitas. The Milpitas facility started manufacturing the v 5.3 systems in February 2013. The serial number is contained in the Product Identification Label (part number Z352548) that is attached to the rear of the console for systems manufactured.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOptiMedica Catalys Precision Laser System; Model No(s). Catalys-U (US), and Catalys-I (Outside US). ProductApril 29, 2015
- FDASTAR Excimer Laser System, Model: STAR S2, Catalog Number: 0030-1479, Product is manufactured and distributedOctober 30, 2013
- FDASTAR ActiveTrak Excimer Laser System, Model: STAR S3, Catalog Number: 0030-2450, Product is manufactured andOctober 30, 2013
- FDASTAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product isOctober 30, 2013
- FDACardio Medical Products, Inc. Defibrillation/Cardioversion/Monophasic or Biphasic Pacing/ECG Electrode Rx OnlyJuly 17, 2013
- FDARigiflexTM 11 Single-Use Achalasia Balloon Dilator, Material Number (UPN) M00554510 Product Usage: The BostonJuly 17, 2013
Questions about this recall
Was Sovereign® Compact Phacoemulsification System, v 5.3, Part Number: CMP680300. The system is used for cataract surgery. recalled?
Yes. FDA published recall Z-1670-2013 on July 17, 2013. Reported hazard: Abbott Medical Optics Inc. (AMO) initiated this Urgent Notice of Medical Device Correction and Advisory Notice to all customers who use the AMO Sovereign® Compact Phacoemulsification System, Version 5.3, because it has the potential to lose the vacuum tone mode during surgery.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail