Back to search
FDAFDA · Food and drugs·Recall number Z-1693-2013

Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models 2008K, 2008K2, 2008K@Home and 2008T

Published July 17, 2013 · Recall initiated: May 24, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Fresenius Medical Care Holdings, Inc.
Distributor
Nationwide Distribution including the states of AR, CA, FL, GA, KY, LA, MA, MD, NC, NJ, NY, PA, PR, SC, TX and VA.
Lot
Part Number: 640919
Risk classification
Class II
Area (FDA)
device
Country
United States