FDAFDA · Food and drugs·Recall number Z-1693-2013

Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models 2008K, 2008K2, 2008K@Home and 2008T

Published July 17, 2013 · Recall initiated: May 24, 2013

Summary

FDA records medical device Class II recall Z-1693-2013, published July 17, 2013 and initiated May 24, 2013. Recalling firm: Fresenius Medical Care Holdings, Inc.; Status: Terminated; Reported quantity: 725 units; Lot: Part Number: 640919. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Fresenius Medical Care Holdings, Inc.
Distributor
Nationwide Distribution including the states of AR, CA, FL, GA, KY, LA, MA, MD, NC, NJ, NY, PA, PR, SC, TX and VA.
Lot
Part Number: 640919
Risk classification
Class II
Area (FDA)
device
Country
United States