Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography

Published July 17, 2013 · Recall initiated: February 28, 2013

Summary

FDA records medical device Class II recall Z-1694-2013, published July 17, 2013 and initiated February 28, 2013. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 3; Lot: Model number 10140000, serial numbers 3118, 3243 and 3309. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
United States Nationwide Distribution in the states of ND, OH and SC.
Lot
Model number 10140000, serial numbers 3118, 3243 and 3309
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography recalled?

Yes. FDA published recall Z-1694-2013 on July 17, 2013. Reported hazard: Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss

What should I do if I have this product?

Voluntary: Firm initiated · Letter