Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography
Published July 17, 2013 · Recall initiated: February 28, 2013
Summary
FDA records medical device Class II recall Z-1694-2013, published July 17, 2013 and initiated February 28, 2013. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 3; Lot: Model number 10140000, serial numbers 3118, 3243 and 3309. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- United States Nationwide Distribution in the states of ND, OH and SC.
- Lot
- Model number 10140000, serial numbers 3118, 3243 and 3309
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAFresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/orJuly 17, 2013
- FDAREF 154200 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size A3 ARCOM UHMWPE / CO-CR-MO Alloy OnlyJuly 17, 2013
Questions about this recall
Was Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography recalled?
Yes. FDA published recall Z-1694-2013 on July 17, 2013. Reported hazard: Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss
What should I do if I have this product?
Voluntary: Firm initiated · Letter