ExacTrac 5.5, an Image Processing System used for patient positioning for radiation therapy or radiosurgery.

Published June 11, 2014 · Recall initiated: April 15, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When using a specific plan update workflow in combination with a non-Brainlab treatment planning system (TPS), there is a possibility in ExacTrac 5.5.5 to generate an unintended blend of two initially separate plans of the same patient.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Brainlab AG
Distributor
Worldwide Distribution - US Nationwide in the states of AZ, AR, CA, FL, KY, MA, NE, NY, TN, TX and the countries of Argentina, Australia, Brazil, Finland, India, Japan, Netherlands, Oman, Russia, and Taiwan
Lot
ExacTrac software version 5.5.5 is included in the following model / catalog numbers: 1) 20818B ET SOFTWARE 5.5.5; 2) 20820 ET SOFTWARE UPDATE 5.0 TO 5.5; 3) 49905D EXACTRAC COMPLETE SYSTEM UPG 3.X TO 5.5; 4) 49908C EXACTRAC SOFTWARE UPGRADE 3.X TO 5.5; 5) 49909A EXACTRAC SOFTWARE UPGRADE 4.X TO 5.5; 6) 49918 ET SOFTWARE UPDATE 5.5.X TO 5.5.5; 7) 49919 ET SOFTWARE UPDATE 5.0.X TO 5.5.5; 8) 49660A ET INFRARED SOFTWARE 5.5.5; 9) 49961 ET DATA PREP/REVIEW SYSTEM; 10) 49986 ET SYSTEM UPGRADE 3.X TO 5.5 NEW VAULT; 11) 49987 ET SYSTEM UPGRADE 4.X TO 5.5 NEW VAULT; 12) 49988 ET SYSTEM UPGRADE 4.X TO 5.5 SAME VAULT; 13) 49989 ET SYSTEM UPGRADE 5.0 TO 5.5 NEW VAULT; 14) 49990 ET SYSTEM UPGRADE 5.0 TO 5.5 SAME VAULT; 15) 49991 ET SYSTEM UPGRADE 5.5 TO 5.5 NEW VAULT
Risk classification
Class II
Area (FDA)
device
Country
Germany