Back to search
FDAFDA · Food and drugs·Recall number Z-1714-2024

ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 16 Holes, 236 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-017-16

Published May 8, 2024 · Recall initiated: March 20, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
Lot
UDI: (01)00889024088559(17)290804(10)64475611 (01)00889024088559(17)290806(10)64475612 Lot Numbers: 64475611 64475612
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.