ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 7-2357-018-04
Published May 8, 2024 · Recall initiated: March 20, 2024
Summary
FDA records medical device Class II recall Z-1715-2024, published May 8, 2024 and initiated March 20, 2024. Recalling firm: Zimmer, Inc.; Status: Ongoing; Lot: UDI: (01)00889024088566(17)290409(10)64325922 (01)00889024088566(17)290414(10)64325923 (01)00889024088566(17)290515(10)64396817 (01)00889024088566(17)290517(10)64396818 (01)00889024088566(17)290831(10)64464677 (01)00889024088566(17)300325(10)64714915 (01)00889024088566(17)300330(10)64714916 (01)00889024088566(17)300321(10)64714917 (01)00889024088566(17)320329(10)65324701 (01)00889024088566(17)320510(10)65387353 (01)00889024088566(17)330220(10)65716463 (01)00889024088566(17)330127(10)65716464 (01)00889024088566(17)330426(10)65982902 (01)00889024088566(17)330426(10)66010705 Lot Numbers: 64325922 64325923 64396817 64396818 64464677 64714915 64714916 64714917 65324701 65387353 65716463 65716464 65982902 66010705. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
- Lot
- UDI: (01)00889024088566(17)290409(10)64325922 (01)00889024088566(17)290414(10)64325923 (01)00889024088566(17)290515(10)64396817 (01)00889024088566(17)290517(10)64396818 (01)00889024088566(17)290831(10)64464677 (01)00889024088566(17)300325(10)64714915 (01)00889024088566(17)300330(10)64714916 (01)00889024088566(17)300321(10)64714917 (01)00889024088566(17)320329(10)65324701 (01)00889024088566(17)320510(10)65387353 (01)00889024088566(17)330220(10)65716463 (01)00889024088566(17)330127(10)65716464 (01)00889024088566(17)330426(10)65982902 (01)00889024088566(17)330426(10)66010705 Lot Numbers: 64325922 64325923 64396817 64396818 64464677 64714915 64714916 64714917 65324701 65387353 65716463 65716464 65982902 66010705
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 16 Holes, 236 MM,May 8, 2024
- FDAZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 6 Holes, 106 MM, Sterile-May 8, 2024
Questions about this recall
Was ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 7-2357-018-04 recalled?
Yes. FDA published recall Z-1715-2024 on May 8, 2024. Reported hazard: Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit